Related Experiment Video
Updated: Jan 16, 2026

Trabecular Meshwork Response to Pressure Elevation in the Living Human Eye
Published on: June 20, 2015
Home Tonometry Diurnal Intraocular Pressure Patterns, Patient Adherence, and Measurement Reliability in a Prospective
Scott W Perkins1, Julia H Joo1, Kevin C Allan2
1Cleveland Clinic Lerner College of Medicine, Case Western Reserve University, Cleveland, OH, USA.
Home monitoring of intraocular pressure (IOP) with the iCare HOME2 tonometer showed good patient adherence and revealed significant diurnal IOP patterns. This device effectively detects out-of-office IOP spikes, correlating well with in-office measurements.
Area of Science:
- Ophthalmology
- Medical Devices
- Patient Monitoring
Background:
- Intraocular pressure (IOP) monitoring is crucial for managing glaucoma.
- Understanding diurnal IOP patterns and out-of-office fluctuations is essential for effective treatment.
- Home-based IOP monitoring offers a potential solution for capturing these variations.
Purpose of the Study:
- To assess patient adherence to home IOP monitoring using the iCare HOME2 tonometer (ICH2).
- To analyze diurnal IOP patterns and their fluctuations over a 24-hour period.
- To evaluate the correlation between home-recorded IOP and in-office tonometry measurements.
Main Methods:
- A prospective cohort study involving 75 patients with open-angle disease.
- Patients used the iCare HOME2 tonometer (ICH2) for self-monitoring, taking 6 measurements daily for 10 days.
- Statistical analysis was performed on adherence data, diurnal IOP patterns, and correlation with in-office IOP readings.
Main Results:
- Good adherence was observed in 65% of patients, with a median of 7.4 measurements/day; however, 19% showed poor adherence.
- Significant diurnal IOP patterns were identified, with a peak in the early morning (3 AM) and a decrease throughout the day.
- Out-of-office IOP variations were significant, with 36% of patients experiencing maximum IOP outside clinic hours.
- Home IOP measurements showed a significant positive correlation with in-office Goldman applanation (r² = 0.35) and iCare IC100 (r² = 0.48) tonometry.
Conclusions:
- The iCare HOME2 tonometer (ICH2) is clinically usable and effective for monitoring intraocular pressure at home.
- The device successfully detects out-of-office IOP spikes and diurnal IOP variations.
- Home IOP monitoring provides valuable data that correlates with traditional in-office measurements.
Related Concept Videos
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Assessment of blood pressure in brachial artery(two-step method)
Errors occurring during blood pressure monitoring
Several factors...
Hypertension III: Clinical Manifestations and Diagnostic Studies
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure