A Method for Quantifying Upper Limb Performance in Daily Life Using Accelerometers
Published on: April 21, 2017
The Concurrent Validity of the ActiGraph LEAP Compared to the ActiGraph GT9X Link for Measuring Physical Activity
Jethro Raphael M Suarez1,2, Joon-Hyuk Park1,3,4, Kworweinski Lafontant2,5
1Department of Mechanical and Aerospace Engineering, University of Central Florida, Orlando, FL, USA.
Background:
The ActiGraph GT9X Link wearable accelerometer has been widely utilized for physical activity (PA) measurement but is no longer in production. The performance of the ActiGraph LEAP, regarded as its successor and released in 2023, remains unevaluated in comparison to other accelerometers. This study aimed to establish concurrent validity of the ActiGraph LEAP device through PA measurements and comparison with the GT9X Link.
Methods:
A total of 26 young adults (mean age = 21 ± 4 years) wore the ActiGraph LEAP and GT9X Link simultaneously for 7 consecutive days on their nondominant wrist. Monitor Independent Movement Summary units were utilized for calculation of daily sedentary time, light PA, and moderate to vigorous PA across valid days of wear (≥10 hr). Nonwear time was determined using a preexisting algorithm. Intraclass correlation coefficient, mean absolute percent error, equivalence tests (with GT9X Link as reference), and Bland-Altman plots assessed agreement between daily values of sedentary time, light PA, and moderate to vigorous PA recorded by each device across all valid days recorded among participants.
Results:
A total of 149 days of data were compared between the two devices. Across all PA levels, acceptable-excellent agreement was found (mean absolute percent error = 3.86%-13.86% and intraclass correlation coefficient = .94-.99), and Bland-Altman plots showed that at least 90% of recorded data fell within the 95% limits of agreement.
Conclusions:
The findings suggest that sedentary time and PA measurements using Monitor Independent Movement Summary units from the ActiGraph LEAP compare well with the GT9X Link, indicating that the LEAP performs similarly to the GT9X Link.