Long-term quality management of third-generation LED light-curing units: A 10-year retrospective analysis
Christian Klein1, Silke Dengel2, Christian Meller1
1Poliklinik für Zahnerhaltung, Universitätsklinikum Tübingen.
Swiss Dental Journal
|March 11, 2026
Summary
Third-generation LED light-curing units (LCUs) demonstrated remarkable irradiance stability over 10 years, even after accidental drops. Routine quality assurance remains crucial despite this device performance.
Area of Science:
- Dental Materials Science
- Biomedical Engineering
Background:
- Radiant exposure from light-curing units (LCUs) is critical for resin-based composite polymerization.
- Effective quality management of light-curing procedures is essential in dental practices.
Purpose of the Study:
- To retrospectively evaluate the performance, service removal frequency, and reasons for third-generation LED LCUs over a decade.
- To assess the long-term irradiance stability and reliability of specific LCU models.
Main Methods:
- Retrospective analysis of 10 years of data from third-generation LED LCUs (Bluephase G2).
- Regular irradiance monitoring using a dental radiometer (Bluephase Meter) at 1-2 month intervals and after incidents.
- Documentation of measurements, incidents, and reasons for removal in practice management software.
Main Results:
- The mean irradiance of 554 measurements was 1,375 ± 68 mW/cm², exceeding the manufacturer's specification.
- No significant loss of tip irradiance was observed even in units monitored for the full 10 years.
- Accidental dropping was the cause in two-thirds of incidents, leading to replacement in about half of those cases.
Conclusions:
- Third-generation LED LCUs exhibit exceptional long-term irradiance stability.
- Despite high performance, continued quality assurance is recommended to account for potential unreported accidents.
Keywords:
Light-curing unitdental officespolymerizationquality managementradiometerresin-based composite