OPTICAL QUALITY IN DECENTRATION OF INTRAOCULAR LENSES: an updated systematic review
Celia Villanueva1, Nuria Garzón1, César Albarrán Diego2
1Department of Optometry and Vision. Faculty of Optics and Optometry. Complutense University Of Madrid, Madrid, Spain.
Topic:
To evaluate the impact of intraocular lens (IOL) decentration on optical quality and performance using optical bench studies and ray-tracing numerical simulations, and to analyze how different IOL designs (monofocal, extended depth of focus [EDoF], multifocal, and toric) respond to decentration.
Clinical Relevance:
IOL decentration may adversely affect visual outcomes, particularly with newer IOL designs. Despite advances in surgical techniques, optimal centration cannot always be guaranteed due to anatomical variability and other factors. Understanding how different IOL designs tolerate decentration is therefore essential for appropriate lens selection and optimized postoperative visual quality.
Methods:
A systematic review of the literature was conducted following PRISMA guidelines. Searches were performed in PubMed, Scopus, and Google Scholar using Boolean operators and truncated terms. Inclusion criteria focused on peer-reviewed optical bench studies published in English over the past 20 years. Data were extracted on optical metrics such as modulation transfer function, Strehl ratio, wavefront aberrations, and point spread function, in relation to various degrees of IOL decentration and lens types.
Results:
A total of 74 studies were included. Aspheric monofocal IOLs with negative spherical aberration showed greater optical degradation with decentration compared to spherical models. Multifocal and EDoF IOLs were more sensitive to decentration, with notable reductions in image quality and contrast sensitivity, especially at higher decentration levels. Toric IOLs were particularly affected when decentration was combined with rotational misalignment. Optical quality loss was directly related to each lens's design and optical profile.
Conclusions:
IOL decentration has a measurable and clinically relevant impact on optical quality, especially in aspheric and multifocal designs. Understanding IOL behavior under decentration conditions is essential for optimizing lens selection and postoperative visual outcomes.
More Related Videos
05:46Author Spotlight: Advancements in Refractive Surgical Correction for Presbyopia and Exploring Postoperative Visual Acuity
Published on: September 20, 2024
04:59Author Spotlight: Enhancing Visual Outcomes in Cataract Surgery: A Novel Technique to Prevent Posterior Capsular Opacification Through IOL Rotation
Published on: July 7, 2023
Related Concept Videos
Focusing of Light in the Eye
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Angle Closure Glaucoma: Treatment
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...