Consistent Quadratic Phase Formation in 3D Fast Spin Echo Using Frequency-Modulated RF Pulses
Naoharu Kobayashi1, Michael Garwood1
1Center for Magnetic Resonance Research, Department of Radiology, University of Minnesota, Minneapolis, Minnesota, USA.
Purpose:
Frequency-modulated (FM) RF pulses achieve broadband excitation with low RF peak power, which is required in MRI with inhomogeneous magnetic fields. However, the quadratic phase generated with FM pulses makes it difficult to use them in fast spin echo (FSE), because even and odd refocused echoes have different spatial phase profiles. In this article, we formulate the condition that can generate a consistent quadratic phase in the two echo components in 3D FSE using FM pulses.
Methods:
The consistent quadratic phase condition in 3D FM-FSE was formulated with the Cayley-Klein parameters. The B1 +-dependent phase in FM excitation was compensated by adjusting the initial phase of FM pulses. The consistent quadratic phase in 3D FM-FSE and the impacts from the B1 +-dependent phase were simulated with extended phase graph (EPG). The EPG simulation was validated by comparing with experimental results at 3 T. Finally, in vivo brain imaging was conducted with T1- and T2-weighted contrasts.
Results:
In EPG simulation, a consistent quadratic phase was generated in all echoes in 3D FM-FSE under the formulated condition. The EPG simulation matched well with experimental results. The B1 +-dependent phase adjustment on FM pulses improved the magnitude profile of the refocused echoes. Despite the presence of large field inhomogeneity, 3D FM-FSE achieved in vivo brain imaging with good T1- and T2-contrasts.
Conclusion:
Achieving consistent quadratic phase in all refocused echoes in 3D FM-FSE makes it possible to use FM pulses in the non-adiabatic regime. 3D FM-FSE is a promising pulse sequence for MRI in inhomogeneous magnetic fields.
Related Concept Videos
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
¹H NMR: Interpreting Distorted and Overlapping Signals
As Δν decreases and the signals move closer, the doublets appear increasingly distorted. The intensities of the inner lines increase at the cost of those of the outer lines as the signals are...
Time and frequency -Domain Interpretation of Phase-lead Control
The design of phase-lead control involves the strategic placement of poles and zeros to balance steady-state error and system...
Atomic Nuclei: Nuclear Relaxation Processes
Two-Dimensional (2D) NMR: Overview
The first step is the preparation period, during which nucleus A is excited with a radiofrequency pulse....