Insertion Torque Characteristics of the KS 3 Implant in Weak Bone, Standardized Extraction-Socket-like, and Maxillary
Na Ri Seo1, Ye-Seul Jung2, Dayeon Park2
1Program in Neuroscience, Department of Dental Science, Graduate School, Seoul National University, Seoul 03080, Republic of Korea.
Bioengineering (Basel, Switzerland)
|June 26, 2026
Summary
The KS 3 implant demonstrated higher insertion torque than the TSIII implant in artificial bone models simulating weak bone, extraction sockets, and the maxillary sinus. These findings highlight differences in implant stability under various simulated clinical conditions.
Area of Science:
- Dental Implantology
- Biomaterials Science
- Surgical Engineering
Background:
- Dental implant stability is crucial for successful osseointegration.
- Insertion torque is a key indicator of primary implant stability.
- Evaluating implant performance in simulated compromised bone conditions is essential.
Purpose of the Study:
- To compare the insertion torque characteristics of the KS 3 implant versus the TSIII implant.
- To assess implant performance in artificial bone models simulating weak bone, extraction sockets, and the maxillary sinus.
- To analyze torque-depth curves and torque-depth integrals.
Main Methods:
- Comparative in vitro study using three artificial bone models: weak bone, extraction socket, and maxillary sinus simulation.
- Measurement of maximum and final insertion torque values.
- Calculation of torque-depth integrals as secondary parameters.
- Statistical analysis using Welch's t-test and two-way ANOVA.
Main Results:
- The KS 3 implant exhibited higher maximum and/or final insertion torque values compared to the TSIII implant across all tested models.
- Directional differences in torque values were observed in the maxillary sinus model based on implant type and residual bone height.
- Torque-depth integral analysis provided supplementary data on cumulative insertion resistance.
Conclusions:
- The KS 3 implant demonstrated superior insertion torque compared to the TSIII implant in the tested artificial bone conditions.
- Findings represent in vitro insertion torque data and do not confirm clinical superiority.
- Observed torque differences in the maxillary sinus model result from combined effects of implant macrodesign and drilling protocols.