Vital sign parameters are not associated with electrical capture during transcutaneous pacing
Joshua M Kimbrell1, Judah A Kreinbrook2, Tanner Smida3
1Department of Pre-Hospital Care, MediSys Health Network, 8900 Van Wyck Expressway, Queens, NY, United States of America; Albert Einstein College of Medicine, Bronx, NY, United States.
Background:
Failure to achieve electrical capture during transcutaneous pacing (TCP) is common and underrecognized. We aimed to identify whether pulse oximetry plethysmograph waveform synchrony, end-tidal capnometry increases, and systolic blood pressure increases were associated with sustained electrical capture during prehospital TCP.
Methods:
This secondary analysis of a multi-center retrospective observational study enrolled patients from four emergency medical service agencies. Sustained electrical capture was defined as a wide QRS and T wave for ≥90% of ongoing TCP time. Pulse oximetry plethysmograph wave synchronization with the TCP rate and vital sign values were obtained from cardiac monitor/defibrillator files. A diagnostic framework was used to evaluate plethysmograph synchronization. We used linear regression modeling to detect whether increases in end-tidal carbon dioxide (ETCO2) and blood pressure measurements were associated with sustained electrical capture.
Results:
Of 253 patients, 29 (11.5%) had sustained electrical capture. An increase in SBP had a sensitivity of 75.0% (42.8, 94.5) and a specificity of 40.0% (29.8, 50.9) for the outcome of sustained electrical capture. An increase in ETCO2 had a sensitivity of 41.7% (15.2, 72.3) and a specificity of 64.4% (53.7, 74.3). Linear regression modeling did not suggest that sustained electrical capture was associated with post-TCP SBP (-2.5 mmHg (-27.3, +22.4)) or post-TCP ETCO2 (2.6 mmHg (-2.9, +8.1)). Plethysmograph synchronization had a sensitivity of 58.8% (32.9, 81.6) and a specificity of 81.1% (73.2, 87.5).
Conclusions:
These findings suggest that plethysmograph waveforms, capnometry, and blood pressure monitoring alone are not reliable indicators of electrical capture during TCP performed by paramedics.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Electrophysiology of Normal Cardiac Rhythm
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...