DDI pacing: indications, expectations, and follow-up
1Division of Cardiology and Cardiovascular Surgery, Toronto General Hospital, Ontario, Canada.
Pacing and Clinical Electrophysiology : PACE
|March 1, 1994
Summary
The DDI pacing mode, offering atrioventricular sequential bradycardia support, was used in 14% of dual chamber pacing patients. Most patients remained in DDI mode, with few reprogramming issues or functional undersensing events.
Area of Science:
- Cardiology
- Biomedical Engineering
- Electrophysiology
Background:
- Dual chamber pacing is crucial for managing bradycardia.
- The DDI pacing mode offers specific advantages for atrioventricular sequential support.
- Patient selection and pacing mode programming are critical for optimal outcomes.
Purpose of the Study:
- To evaluate the efficacy and safety of the DDI pacing mode in patients requiring dual chamber pacing.
- To assess the long-term performance and reprogramming outcomes of DDI pacing.
- To identify potential complications and management strategies associated with DDI pacing.
Main Methods:
- Retrospective analysis of 65 patients programmed to DDI mode between February 1985 and March 1990.
- Utilized Pacesetter 283 or 285 pulse generators programmed to DDI.
- Follow-up included predischarge, 6 weeks, 3 months, and 6 months post-discharge, with a mean follow-up of 31 months.
Main Results:
- 83% of patients remained in the DDI mode throughout the follow-up period.
- Six patients experienced effective VVI pacing due to ventriculoatrial conduction with atrial output inhibition.
- Four patients had functional undersensing, resolved by reprogramming the ventricular blanking period to 13 msec.
Conclusions:
- The DDI pacing mode is a viable option for dual chamber pacing, particularly for patients with sick sinus syndrome and AV block.
- Most patients tolerate DDI pacing well, with a low rate of reprogramming and manageable functional undersensing.
- Careful programming of the ventricular blanking period can effectively address functional undersensing issues in DDI mode.