External Excitation of Neurons Using Electric and Magnetic Fields in One- and Two-dimensional Cultures
Published on: May 7, 2017
Three-dimensional current density distribution under surface stimulation electrodes
A M Sagi-Dolev1, D Prutchi, R H Nathan
1Biomedical Engineering Program, Ben-Gurion University of the Negev, Beer-Sheva, Israel.
Controlling current density distribution under functional neuromuscular stimulation (FNS) electrodes minimizes overflow to non-target tissues. This study presents a novel method using a phantom model and advanced circuits to map 3-D current density for optimized FNS.
Area of Science:
- Biomedical Engineering
- Neuroscience
- Electrical Engineering
Background:
- Functional neuromuscular stimulation (FNS) can cause overflow to non-target tissues.
- Controlling current density distribution under surface electrodes is key to reducing this overflow.
- Optimizing FNS parameters is crucial for effective muscle activation and patient comfort.
Purpose of the Study:
- To introduce a method for acquiring 3-D current density distributions under complex FNS electrode geometries.
- To develop a system simulating excitable tissue response to external stimulation.
- To enable characterization of stimulation parameters for targeted muscle activation with minimal overflow.
Main Methods:
- Utilized a phantom model with a skin impedance layer simulating FNS parameters.
- Developed signal acquisition and processing circuits to mimic tissue response.
- Introduced a data analysis method for characterizing stimulation intensity, electrode geometry, and pulse waveform.
Main Results:
- Successfully acquired 3-D current density distributions under various electrode geometries.
- Presented results as 3-D attenuation coefficient maps.
- Demonstrated the method's applicability in optimizing FNS for targeted muscle activation and reduced side effects.
Conclusions:
- The developed method effectively maps 3-D current density under FNS electrodes.
- This approach allows for precise control of stimulation parameters to minimize overflow and discomfort.
- Optimized FNS delivery can enhance therapeutic outcomes and patient experience.
More Related Videos
Related Concept Videos
Equipotential Surfaces and Conductors
Current Density
Boundary Conditions for Current Density
Ampere-Maxwell's Law: Problem-Solving
To solve the problem, we can use the equations from the analysis of an RC circuit and Maxwell's version of Ampère's law.
For the first part of the problem,...
The Electrical Double Layer