Plasma microdialysis. A technique for continuous plasma sampling in freely moving rats
1Department of Physiology, Southern Illinois University, School of Medicine, Carbondale 62901-6512.
Journal of Pharmacological and Toxicological Methods
|April 1, 1993
Summary
This study demonstrates a novel microdialysis probe for continuous plasma sampling in rats. Essential for accurate glucose monitoring, continuous heparinization and in vivo calibration are crucial for successful plasma microdialysis.
Area of Science:
- Physiology
- Biomedical Engineering
- Analytical Chemistry
Background:
- Continuous plasma sampling is vital for real-time physiological monitoring.
- Traditional methods like repeated blood draws are invasive and stressful.
- Developing minimally invasive techniques for chronic analyte measurement is a key research area.
Purpose of the Study:
- To develop and validate a specialized microdialysis probe for continuous plasma sampling in freely moving rats.
- To assess the probe's efficacy in measuring glucose, protein, and luteinizing hormone.
- To determine the necessity of heparinization and in vivo calibration for accurate results.
Main Methods:
- A custom-designed microdialysis probe with side-arm tubing was utilized.
- The probe was inserted into the right atrium of freely moving rats.
- In vivo probe calibration and continuous heparin infusion were performed.
- Plasma glucose, protein, and luteinizing hormone concentrations were measured.
Main Results:
- Continuous heparinization maintained a stable glucose recovery rate (18%) over 24 hours.
- Non-heparinized rats showed a rapid drop in glucose recovery to 1%.
- In vivo calibration was essential for accurate plasma glucose concentration determination.
- Plasma protein recovery was low (0.07% initially, declining to 0.03%) and luteinizing hormone was undetectable.
Conclusions:
- Continuous heparinization is essential to prevent clotting and ensure probe function.
- In vivo probe calibration is critical for accurate quantification of plasma analytes.
- The developed microdialysis system is suitable for sampling non-protein-bound molecules in plasma.
- Further optimization may be needed for protein and peptide hormone measurements.