Related Experiment Video
Updated: Aug 9, 2026

Post Column Derivatization Using Reaction Flow High Performance Liquid Chromatography Columns
Published on: April 26, 2016
Homocysteine and other thiols determined in plasma by HPLC and thiol-specific postcolumn derivatization
A Andersson1, A Isaksson, L Brattström
1Department of Clinical Chemistry, University Hospital, Lund, Sweden.
Abstract:
We describe a versatile high-performance liquid-chromatographic method for determining homocysteine and other plasma sulfhydryls. Using three different procedures for preparation of plasma, we determined total, free (non-protein-bound), and reduced forms of homocysteine, cysteine, glutathione, cysteinylglycine, and gamma-glutamylcysteine in human plasma. Sample preparation involves disulfide reduction with dithiothreitol and protein precipitation with sulfosalicylic acid. The assay utilizes isocratic reversed-phase ion-pair liquid chromatography at pH 2.4, postcolumn derivatization with 4,4'-dithiodipyridine, and colorimetric detection at 324 nm. The intra-assay precision (CV) of the method for total homocysteine is 1.5%; the interassay precision over 2.5 months is 2.5%. The detection limit for homocysteine is < 50 nmol/L plasma.
More Related Videos
10:09A Straightforward Method for Glucosinolate Extraction and Analysis with High-pressure Liquid Chromatography (HPLC)
Published on: March 15, 2017
07:16Resin-Assisted Capture Coupled with Isobaric Tandem Mass Tag Labeling for Multiplexed Quantification of Protein Thiol Oxidation
Published on: June 21, 2021