Reticulocyte quantification by Coulter MAXM VCS (volume, conductivity, light scatter) technology
M Buttarello1, P Bulian, M D Prà
1Laboratorio Analisi Chimico-Cliniche e Microbiologiche, Ospedale Geriatrico, Padova, Italia.
Clinical Chemistry
|December 1, 1996
Summary
The Coulter MAXM analyzer accurately quantifies reticulocytes, showing good correlation with manual methods and other automated analyzers. It offers improved precision over manual counting, though some concentration-dependent biases were observed.
Area of Science:
- Hematology
- Clinical Chemistry
- Laboratory Medicine
Background:
- Reticulocyte counts are crucial indicators of bone marrow erythropoietic activity.
- Accurate reticulocyte quantification is essential for diagnosing and monitoring various anemias and hematological disorders.
- Evaluating new automated analyzers is vital for improving laboratory efficiency and diagnostic accuracy.
Purpose of the Study:
- To assess the analytical performance of the Coulter MAXM analyzer for reticulocyte quantification.
- To compare the MAXM analyzer's results with a manual microscopic method and the Sysmex R-1000 automated analyzer.
- To determine the clinical utility of the MAXM analyzer in a routine laboratory setting.
Main Methods:
- Analysis of 330 patient samples using the Coulter MAXM analyzer.
- Comparison with reticulocyte counts obtained by a manual microscopic method (NCCLS H44-P).
- Comparison with results from the Sysmex R-1000 automated analyzer.
Main Results:
- The Coulter MAXM analyzer demonstrated lower imprecision (CV) compared to the manual microscopic method across all concentration levels.
- Good overall correlation was observed between the MAXM analyzer and the manual method (r2 = 0.87).
- The MAXM analyzer showed concentration-dependent biases, overestimating at low counts and underestimating at normal and high counts when compared to the manual method.
Conclusions:
- The Coulter MAXM analyzer is a precise automated method for reticulocyte counting, outperforming manual methods in imprecision.
- While showing good overall agreement, users should be aware of potential biases at different reticulocyte concentrations.
- The MAXM analyzer provides satisfactory sensitivity for high reticulocyte counts, making it a viable option for clinical laboratories.