[A clinical comparative study of tyndallometry using 2 laser flare meters]
N X Nguyen1, P Martus, M Küchle
1Augenklinik mit Poliklinik, Universität Erlangen-Nürnberg.
Summary
The Kowa FM-500 and FC-1000 laser flare meters show comparable results for measuring aqueous flare in normal eyes and those with mild to moderate uveitis. However, significant differences may occur with very high flare readings.
Area of Science:
- Ophthalmology
- Biomedical Optics
- Medical Instrumentation
Background:
- The Kowa laser flare meter FM-500 is a simplified version of the FC-1000.
- Both instruments utilize lasers to measure aqueous flare, a marker of intraocular inflammation.
Purpose of the Study:
- To compare flare measurements between the Kowa FM-500 and the established Kowa FC-1000 laser flare-cell meter.
- To assess the in vivo reproducibility of both instruments for quantifying aqueous flare.
Main Methods:
- Compared flare measurements from the FM-500 and FC-1000 in 10 normal subjects and 21 uveitis patients.
- Examinations were performed by a single examiner under comparable, randomized conditions.
- Reproducibility was assessed by repeat measurements on normal eyes the following day; statistical analysis included linear regression and intraclass correlation.
Main Results:
- Flare values did not significantly differ between the FM-500 and FC-1000 in normal eyes (3.3 vs 3.5 photon counts/ms) or uveitis eyes (16.2 vs 16.4 photon counts/ms).
- High correlations were observed between instruments in normal eyes (r=0.73) and very high correlations in eyes with mild to moderate flare (R=0.99).
- Intraclass correlation coefficients were high for both instruments (0.70-0.76 in normal eyes, 0.99 in uveitis eyes), though the FM-500 indicated higher values in one case of acute fibrinous iridocyclitis with very high flare.
Conclusions:
- The Kowa FM-500 and FC-1000 provide comparable aqueous flare measurements in eyes without or with mild to moderate blood-aqueous barrier breakdown.
- Discrepancies in measurements may arise when assessing eyes with significantly elevated flare levels.