Attenuation characteristics and application of gadolinium-DTPA in fast helical computed tomography
W Luboldt1, M De Santis, A von Smekal
1Department of Radiology, University Hospital Grosshadern, Munich, Germany.
Rationale And Objectives:
The computed tomography (CT) attenuation characteristics of gadolinium (Gd)-DTPA have been analyzed both in vitro and in vivo to evaluate the possibility of its substitution for iodine in patients with contraindications for iodine utilization.
Methods:
In vitro attenuation values of Gd-DTPA were measured relative to those of iodine under variation of concentration, voltage, preattenuation, and beam hardening. In vivo iodine attenuation values were obtained for different injection rates (3.0 mL/second and 4.5 mL/second) to calculate the attenuation values of 0.5 molar Gd-DTPA to be expected in vivo by the help of the previously obtained attenuation-concentration curves. Subsequently, CT enhancement after an injection of 0.3 mmol/kg body weight Gd-DTPA was evaluated in five patients with contraindications for iodine administration.
Results:
The injection of a 0.7 molar Gd-DTPA solution at a flow rate of 4.5 mL/second was calculated to be equivalent to that of 2.36 molar iodine at a flow rate of 3 mL/ second. Sufficient enhancement for diagnosis of dissections/ aneurysms of arterial vessels could be achieved. Both arterial and venous cervical or thoracic vessels could be enhanced to 100 Hounsfield units with a bolus injection. In contrast to the thyroid gland, the parenchyma of the liver and the kidney was not sufficiently enhanced.
Conclusions:
Gadolinium-DTPA is more effective than iodine in x-ray attenuation; however, its present concentration of 0.5 molar and permitted dose of 0.3 mmol/kg body weight are insufficient to provide parenchymal enhancement with CT. The visualization of arterial vessels appears possible with gadolinium enhanced CT but requires optimal timing and fast scanning capabilities.