Related Experiment Video
Updated: Aug 29, 2026

Semi-Targeted Ultra-High-Performance Chromatography Coupled to Mass Spectrometry Analysis of Phenolic Metabolites in Plasma of Elderly Adults
Published on: April 22, 2022
A semi-micromethod for determination of oxalate in human plasma
Tadeusz Porowski1, Władysław Gałasiński
1Clinic of Children Diseases Medical Academy, 15-274 Bialystok, 17 Waszyngtona Str., Poland. tporowski@pb.bialystok.pl
Abstract:
An enzymatic semi-micromethod for oxalate determination in human plasma was elaborated. The principle of the method depends on the oxalate isolation from deproteinized plasma, following determination by the calorimetric oxalate oxidase-peroxidase-indamine system. This method protects against internal oxalate losses and excludes an interference of contaminations. Results, obtained by this method, were reliable and ideally suited for use as real normal values (less than or equal to 6 microM) of oxalate content in the plasma of healthy individuals. The elaborated method, which can assay plasma oxalate accurately in normal individuals as well as in hyperoxalemic conditions is superior to those previously used. The procedure of semi-micromethod does not require expensive equipments and apparatus: it is simple and easy to perform in every laboratory and takes little time.