Accuracy of a manual torque application device for morse-taper implants: a technical note.
Murat C Cehreli1, Kivang Akça, Ergin Tönük
1Department of Prosthodontics, Faculty of Dentistry, Hacettepe University, Ankara, Turkey. mcehreli@hotmail.com
The International Journal of Oral & Maxillofacial Implants
|October 29, 2004
Summary
New and used manual torque devices for Morse taper implants were compared. Used devices showed a slight decrease in torque output, with heavily used devices applying approximately 1-1.5 Ncm less torque.
Area of Science:
- Dental Implantology
- Biomaterials Science
- Mechanical Engineering
Background:
- Accurate torque application is crucial for dental implant stability and long-term success.
- Manual torque devices are commonly used in implant dentistry, but their performance over time is not fully understood.
Purpose of the Study:
- To evaluate and compare the torque output of new versus used manual torque devices specifically designed for Morse taper implants.
- To determine if clinical usage affects the accuracy and consistency of torque delivery.
Main Methods:
- Fifteen ITI manual torque devices were categorized into three groups based on usage: new, moderately used (50-200 times), and heavily used (500-1,000 times).
- Torque measurements were performed using a custom wrench with strain gauges connected to a data acquisition system at 35 Ncm and 15 Ncm target settings.
- Torque data were collected at 10,000 Hz and converted to torque units for analysis.
Main Results:
- New manual torque devices delivered significantly higher torque compared to used devices at the 35 Ncm target (P < .05).
- Heavily used devices (500-1,000 uses) exhibited a torque output approximately 1.5 Ncm lower at 35 Ncm and 1 Ncm lower at 15 Ncm compared to newer devices.
- A slight but measurable decrease in torque output was observed with increased clinical use.
Conclusions:
- ITI manual torque devices demonstrate consistent performance but experience a slight reduction in torque output with cumulative clinical usage.
- Regular calibration or replacement of manual torque devices may be necessary to maintain optimal torque accuracy in clinical practice.