Laboratory assessment of oxygenation in methemoglobinemia
Shannon Haymond1, Rohit Cariappa, Charles S Eby
1Department of Pathology and Immunology, Washington University School of Medicine, 660 South Euclid Ave., St. Louis, MO 63110, USA.
Background:
This case conference reviews laboratory methods for assessing oxygenation status: arterial blood gases, pulse oximetry, and CO-oximetry. Caveats of these measurements are discussed in the context of two methemoglobinemia cases.
Cases:
Case 1 is a woman who presented with increased shortness of breath, productive cough, chest pain, nausea, fever, and cyanosis. CO-oximetry indicated a carboxyhemoglobin (COHb) fraction of 24.9%. She was unresponsive to O(2) therapy, and no source of carbon monoxide could be noted. Case 2 is a man who presented with syncope, chest tightness, and signs of cyanosis. His arterial blood was dark brown, and CO-oximetry showed a methemoglobin (MetHb) fraction of 23%.
Issues:
Oxygen saturation (So(2)) can be measured by three approaches that are often used interchangeably, although the measured systems are quite different. Pulse oximetry is a noninvasive, spectrophotometric method to determine arterial oxygen saturation (S(a)O(2)). CO-oximetry is a more complex and reliable method that measures the concentration of hemoglobin derivatives in the blood from which various quantities such as hemoglobin derivative fractions, total hemoglobin, and saturation are calculated. Blood gas instruments calculate the estimated O(2) saturation from empirical equations using pH and Po(2) values. In most patients, the results from these methods will be virtually identical, but in cases of increased dyshemoglobin fractions, including methemoglobinemia, it is crucial that the distinctions and limitations of these methods be understood.
Conclusions:
So(2) calculated from pH and Po(2) should be interpreted with caution as the algorithms used assume normal O(2) affinity, normal 2,3-diphosphoglycerate concentrations, and no dyshemoglobins or hemoglobinopathies. CO-oximeter reports should include the dyshemoglobin fractions in addition to the oxyhemoglobin fraction. In cases of increased MetHb fraction, pulse oximeter values trend toward 85%, underestimating the actual oxygen saturation. Hemoglobin M variants may yield normal MetHb and increased COHb or sulfhemoglobin fractions measured by CO-oximetry.
More Related Videos
09:04Skeletal Muscle Neurovascular Coupling, Oxidative Capacity, and Microvascular Function with 'One Stop Shop' Near-infrared Spectroscopy
Published on: February 20, 2018
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Oxygen Transport in the Blood
Hypoxia
Types of Hypoxia
There are four primary types of hypoxia, each resulting from a different cause:
1. Anemic hypoxia: This type occurs due to insufficient oxygen delivery caused by a lack of red blood cells (RBCs) or RBCs with abnormal or...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...