Resolving for changes in concentration of tissue chromophores in the human adult brain with near-infrared spectroscopy has generally been based on the assumption that optical scattering and pathlength remain constant. We have used a novel hybrid optical spectrometer that combines multi-distance frequency and broadband systems to investigate the changes in scattering and pathlength during a Valsalva manoeuvre in 8 adult volunteers. Results show a significant increase in the reduced scattering coefficient of 17% at 790nm and 850nm in 4 volunteers during the peak of the Valsalva. However, these scattering changes do not appear to significantly affect the differential pathlength factor and the tissue haemoglobin concentration measurements.