1Department of Cardiology, Royal Victoria Hospital, Belfast, UK.
Background:
Conductor cable externalization with protrusion (CCE*) is highly prevalent among the Riata 8F and ST 7F defibrillation (DF) leads and infrequently present in the QuickSite and the QuickFlex coronary sinus (CS) leads (St. Jude Medical, Sylmar, CA, USA). A model for CCE* based on differential lead component pulling and conjugate extension with reciprocal compression-bending was developed. Extension of a proximal lead body segment by pectoral or cardiac movements causes reciprocal compression-bending of a distal lead body segment mediated by inextensible conductor cables running down a lead body fixed at various points by fibrous adhesions. The "sawing" action of these cables under tension causes inside-out abrasion of insulation leading to CCE*.
Methods:
DF leads from different manufacturers and the QuickFlex and QuickFlex μ CS leads were subjected to simulated differential pulling.
Results:
Restitution from differential pulling followed three patterns: complete, partial without escalation, and incomplete with escalation. Only the last pattern (only shown by the Riata 8F and ST 7F leads) was associated with an increased risk to CCE*. For CS leads, deformation concentrated on the more flexible segment when the lead body did not have a uniform construction.
Conclusions:
The Durata, Riata ST Optim, QuickFlex μ, and Quartet leads should be relatively immune to CCE*. The Durata leads are extremely resistant to longitudinal deformation and probably cause mediastinal displacement rather than differential pulling in response to pectoral movements in vivo. Implantation techniques and lead designs can be used to minimize the risk of CCE*. A bench test for CCE* can be constructed.
More Related Videos
06:17Quantifying the Relative Thickness of Conductive Ferromagnetic Materials Using Detector Coil-Based Pulsed Eddy Current Sensors
Published on: January 16, 2020
09:26In Situ Time-dependent Dielectric Breakdown in the Transmission Electron Microscope: A Possibility to Understand the Failure Mechanism in Microelectronic Devices
Published on: June 26, 2015
Related Concept Videos
Charging Conductors By Induction
Generally, conductors like metals do not allow any excess charge to be present on them. Any excess charge added to metals easily flows away, for example, when a metal is placed on the Earth. This process is called earthing.
However, conductors can be charged by a process called induction. For example, consider charging a...
Equipotential Surfaces and Conductors
Resistance and Conductance
Various factors impact the resistance of a conductor. Spiraling in stranded conductors increases their length...
Electric Field Inside a Conductor
Suppose a piece of metal is placed near a positive charge. The free electrons in the metal are attracted to the external positive charge and migrate freely toward that region. This region then has...
Cable Subjected to a Distributed Load
The Electrical Double Layer