Interventional device visualization with toroidal transceiver and optically coupled current sensor for radiofrequency
Maryam Etezadi-Amoli1, Pascal Stang, Adam Kerr
1Magnetic Resonance Systems Research Laboratory, Department of Electrical Engineering, Stanford University, Stanford, California, USA.
Purpose:
The development of catheters and guidewires that are safe from radiofrequency (RF) -induced heating and clearly visible against background tissue is a major challenge in interventional MRI. An interventional imaging approach using a toroidal transmit-receive (transceive) coil is presented. This toroidal transceiver allows controlled, low levels of RF current to flow in the catheter/guidewire for visualization, and can be used with conductive interventional devices that have a localized low-impedance tip contact.
Methods:
Toroidal transceivers were built, and phantom experiments were performed to quantify transmit power levels required for device visibility and to detect heating hazards. Imaging experiments in a pig cadaver tested the extendibility to higher field strength and nonphantom settings. A photonically powered optically coupled toroidal current sensor for monitoring induced RF currents was built, calibrated, and tested using an independent image-based current estimation method.
Results:
Results indicate that high signal-to-noise ratio visualization is achievable using milliwatts of transmit power-power levels orders of magnitude lower than levels that induce measurable heating in phantom tests. Agreement between image-based current estimates and RF current sensor measurements validates sensor accuracy.
Conclusion:
The toroidal transceiver, integrated with power and current sensing, could offer a promising platform for safe and effective interventional device visualization.
Related Concept Videos
Torque On A Current Loop In A Magnetic Field
Consider a rectangular current-carrying loop containing N turns of wire, placed in a uniform magnetic field. The net force on a current-carrying loop...
Magnetic Field Of A Current Loop
Toroids
When connected to a supply, the magnetic field generated in the toroid has field lines circular and concentric to its axis. Conventionally, the direction of this magnetic field is expressed using the right-hand rule. If the fingers of the right hand curl in the current direction, the thumb points in the...
Controlled-Current Coulometry: Overview
Force On A Current Loop In A Magnetic Field
Applications of EMF Measurements