Direct comparison between cerebral oximetry by INVOS(TM) and EQUANOX(TM) during cardiac surgery: a pilot study
A Pisano1, N Galdieri1, T P Iovino1
1Cardiac Anesthesia and Intensive Care Unit, "Monaldi" Hospital A.O.R.N. "Dei Colli", Naples, Italy.
Introduction:
Several near-infrared spectroscopy oximeters are commercially available for clinical use, with lack of standardization among them. Accordingly, cerebral oxygen saturation thresholds for hypoxia/ischemia identified in studies conducted with INVOS(TM) models do not necessarily apply to other devices. In this study, the measurements made with both INVOS(TM) and EQUANOX(TM) oximeters on the forehead of 10 patients during conventional cardiac surgery are directly compared, in order to evaluate the interchangeability of these two devices in clinical practice.
Methods:
Cerebral oxygen saturation measurements were collected from both INVOS(TM) 5100C and EQUANOX(TM) 7600 before anesthetic induction (baseline), two minutes after tracheal intubation, at cardiopulmonary bypass onset/offset, at aortic cross-clamping/unclamping, at the end of surgery and whenever at least one of the two devices measured a reduction in cerebral oxygen saturation equal to or greater than 20% of the baseline value. Bland-Altman analysis was used to compare the bias and limits of agreement between the two devices.
Results:
A total of 140 paired measurements were recorded. The mean bias between INVOS(TM) and EQUANOX(TM) was -5.1%, and limits of agreement were ±16.37%. Considering the values as percent of baseline, the mean bias was -1.43% and limits of agreement were ±16.47. A proportional bias was observed for both absolute values and changes from baseline.
Conclusions:
INVOS(TM) and EQUANOX(TM) do not seem to be interchangeable in measuring both absolute values and dynamic changes of cerebral oxygen saturation during cardiac surgery. Large investigations, with appropriate design, are needed in order to identify any device-specific threshold.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...