Related Experiment Video
Updated: Mar 29, 2026

05:19
Author Spotlight: Advancing Research in Microbial Autoaggregation Using Imaging Flow Cytometry
Published on: September 29, 2023
1.4K
Measurement of Microsphere Concentration Using a Flow Cytometer with Volumetric Sample Delivery
Lili Wang1, Yu-Zhong Zhang2, Steven Choquette1
1National Institute of Standards and Technology, Gaithersburg, MD 20899.
Summary
Accurate microsphere concentration measurement is crucial for quantitative flow cytometry. A new method using a syringe-based system, accounting for volume bias and optimal buffer conditions, achieves ~4% uncertainty for reliable microsphere quantification.
Area of Science:
- Quantitative Flow Cytometry
- Microsphere Characterization
- Biophysical Measurement Techniques
Background:
- Accurate microsphere concentration determination is essential for assigning Equivalent Reference Fluorophores (ERF) units in quantitative flow cytometry.
- Existing methods may lack precision due to variations in sample delivery and buffer composition.
Purpose of the Study:
- To evaluate a syringe-based sample delivery system for precise microsphere concentration measurement.
- To establish an optimized protocol for accurate microsphere quantification using flow cytometry.
Main Methods:
- Calibrated syringe volume delivery and draw modes to identify systematic volume-dependent bias.
- Measured microsphere concentrations using lyophilized microspheres and fluorescein microspheres in various buffer solutions (PBS, PBS+BSA, PBS+TWEEN 20).
- Developed a corrected volume calculation by subtracting a 13 µL offset derived from calibration.
Main Results:
- Syringe volume calibration revealed a -13 µL offset and a systematic volume-dependent bias (slope 0.93).
- Phosphate-buffered saline with 0.05% bovine serum albumin (PBS+BSA) buffer yielded the most consistent count standard concentration.
- The optimized procedure, using PBS+BSA and a corrected delivered volume, yielded results consistent with certified values and relative measurements, with ~4% uncertainty.
Conclusions:
- A syringe-based flow cytometer system can accurately measure microsphere concentrations when accounting for volume bias.
- Using PBS+BSA buffer and a corrected delivered volume (indicated volume - 13 µL) is recommended for optimal microsphere quantification.
- The developed method provides a reliable and accurate approach for microsphere concentration determination in flow cytometry applications.