Hybrid Closed-Loop Insulin Delivery in Type 1 Diabetes During Supervised Outpatient Conditions.
Benyamin Grosman1, Jacob Ilany2, Anirban Roy3
1Medtronic MiniMed, Northridge, CA, USA benyamin.grosman@medtronic.com.
The Medtronic Hybrid Closed-Loop (HCL) system demonstrated safety and efficacy in type 1 diabetes management. Personalized HCL parameters improved glucose control, showing more time in range and less time below range compared to default settings.
Area of Science:
- Endocrinology
- Biomedical Engineering
- Diabetes Technology
Background:
- Type 1 diabetes management requires continuous glucose monitoring and insulin delivery.
- Hybrid Closed-Loop (HCL) systems represent an advancement in automated insulin delivery.
- Evaluating the efficacy and safety of novel HCL systems is crucial for clinical adoption.
Purpose of the Study:
- To assess the efficacy and safety of the Medtronic Hybrid Closed-Loop (HCL) system in individuals with type 1 diabetes.
- To compare the performance of HCL system using default versus personalized parameters.
- To investigate the impact of meal carbohydrate estimation accuracy on glycemic control.
Main Methods:
- A supervised outpatient study involving nine subjects with type 1 diabetes.
- Utilized a prototype HCL system with a sensor-augmented insulin pump and an Android-based control algorithm.
- Compared closed-loop control using default and personalized parameters, with meal carbohydrate intake estimated via metabolic kitchen, dietician, or subject estimates.
Main Results:
- The HCL system achieved an overall 80% time in range (70-180 mg/dL) with minimal hypoglycemia (0.79% time <70 mg/dL).
- Personalized HCL parameters resulted in significantly improved glucose control (137 ± 37 mg/dL mean glucose) compared to default parameters (158 ± 49 mg/dL mean glucose).
- Meal carbohydrate estimation method did not significantly impact postprandial glycemia, with most subjects underestimating intake.
Conclusions:
- The Medtronic HCL system is both efficacious and safe for managing type 1 diabetes.
- Personalized HCL parameters enhance glycemic control, leading to increased time in range and reduced time below range.
- Accurate estimation of meal carbohydrates did not appear to be a critical factor for improving postprandial glycemic control in this study.
More Related Videos
Related Concept Videos
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Insulin Formulations: Types and Delivery
Short-acting insulins are divided into...
Diabetes Mellitus: Overview and Type I Subtype
Type 1 diabetes is an autoimmune disease in which the immune system mistakenly attacks and destroys the insulin-producing beta cells in the pancreas. As a result, the body is unable to produce sufficient insulin, and individuals with...
Diabetes: Management and Pharmacotherapy
Insulin remains the cornerstone of treatment for most patients with type 1 and many...
Diabetes Mellitus: Type 2 and Gestational
Insulin: Biosynthesis, Chemistry, and Preparation
Damage or functional impairment of β-cells inhibits insulin production, leading to diabetes. Diabetes treatment...