Comparison of clinical MRI liver iron content measurements using signal intensity ratios, R 2 and R 2
Jurgen H Runge1, Erik M Akkerman2, Marian A Troelstra2
1Department of Radiology, Academic Medical Center, University of Amsterdam, Meibergdreef 9, 1105AZ, Amsterdam, The Netherlands. j.h.runge@amc.uva.nl.
Signal intensity ratio (SIR) and R2* MRI methods offer higher success rates for liver iron content (LIC) measurement than R2. R2* is favored for its speed and broad measurement range.
Area of Science:
- Radiology
- Medical Imaging
- Hepatology
Background:
- Accurate liver iron content (LIC) measurement is crucial for managing iron overload disorders.
- Magnetic Resonance Imaging (MRI) offers non-invasive methods for LIC assessment.
- Routine clinical practice involves various MRI techniques for LIC quantification.
Purpose of the Study:
- To compare the clinical utility of three MRI-based liver iron content (LIC) measurement techniques.
- To evaluate the success rates and diagnostic accuracy of signal intensity ratio (SIR), R2, and R2* methods.
- To determine the most suitable MRI method for routine LIC assessment over a 6-year period.
Main Methods:
- Retrospective analysis of 114 patients undergoing 1.5T MRI for LIC measurement (2008-2013).
- Comparison of signal intensity ratio (SIR), MR-relaxometry (R2 and R2*) methods for LIC quantification.
- Assessment of success rates, interobserver agreement (ICC), and diagnostic accuracy (ROC analysis) against elevated R2*.
Main Results:
- Successful LIC data acquisition rates were 89% for SIR, 62% for R2, and 98% for R2*.
- All methods demonstrated excellent interobserver agreement (ICC > 0.997).
- R2 and serum ferritin showed high diagnostic accuracy in detecting iron overload indicated by elevated R2*.
Conclusions:
- Signal intensity ratio (SIR) and R2* methods are preferable to R2 due to higher success rates in routine clinical practice.
- R2* is favored over SIR for MRI-based LIC measurement owing to its shorter acquisition time and wider measurable LIC range.
- MRI techniques, particularly R2*, provide reliable and efficient assessment of liver iron overload.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies IV: Magnetic Resonance Imaging
Calibration Curves: Correlation Coefficient
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...