Related Experiment Video
Updated: Mar 11, 2026

High Throughput Microfluidic Rapid and Low Cost Prototyping Packaging Methods
Published on: December 23, 2013
A GENERIC PACKAGING TECHNIQUE USING FLUIDIC ISOLATION FOR LOW-DRIFT IMPLANTABLE PRESSURE SENSORS
A Kim1, C R Powell2, B Ziaie1
1Birck Nanotechnology Center, Purdue University, West Lafayette, IN, USA.
Abstract:
This paper reports on a generic packaging method for reducing drift in implantable pressure sensors. The described technique uses fluidic isolation by encasing the pressure sensor in a liquid-filled medical-grade polyurethane balloon; thus, isolating it from surrounding aqueous environment that is the major source of baseline drift. In-vitro tests using commercial micromachined piezoresistive pressure sensors show an average baseline drift of 0.006 cmH2O/day (0.13 mmHg/month) for over 100 days of saline soak test, as compared to 0.101 cmH2O/day (2.23 mmHg/month) for a non-fluidic-isolated one soaked for 18 days. To our knowledge, this is the lowest reported drift for an implantable pressure sensor.