Rapid focus map surveying for whole slide imaging with continuous sample motion.
Optics Letters
|September 29, 2017
Summary
A new method rapidly surveys focus maps for whole slide imaging (WSI) without axial scanning. This technique enables faster, more accurate digital pathology slide scanning and analysis.
Area of Science:
- Digital Pathology
- Microscopy Imaging
Background:
- Whole slide imaging (WSI) is cleared for primary diagnosis, but high-speed accurate focusing remains a challenge.
- Traditional WSI systems require static samples and axial scanning for focus mapping, limiting speed and efficiency.
Purpose of the Study:
- To develop a novel, high-speed focus map surveying method for WSI.
- To overcome the limitations of traditional focus mapping techniques in WSI.
Main Methods:
- Illuminating samples with two LEDs and recovering focus points using 1D autocorrelation analysis.
- Implementing a method that requires no axial scanning, additional camera, or lens.
- Enabling continuous sample motion during the focus surveying process.
Main Results:
- Demonstrated mean focusing errors of approximately 0.08 μm in static mode and 0.17 μm in continuous motion mode with a 20× objective lens.
- The method is robust for both stained and transparent samples.
- Achieved high-speed, accurate focus mapping without complex hardware.
Conclusions:
- The novel focus map surveying method offers a simple, robust, and accurate solution for WSI systems.
- This technique has the potential to standardize WSI imaging performance and benefit digital pathology.
- Applications extend to high-content microscopy, including time-lapse live-cell imaging.
Related Concept Videos
Confocal Fluorescence Microscopy
21.4K
Confocal microscopy is an advanced microscopic technique. The prime advantage of the confocal microscope over other microscopy techniques is its ability to block the out-of-focus light from the illuminated samples using pinholes. It is widely used with fluorescence optics to obtain high-resolution, sharp contrast images. Unlike optical microscopes, confocal microscopes use a focused beam of light laser to scan the entire sample surface at different z-planes. These microscopes are, therefore,...
21.4K
Super-resolution Fluorescence Microscopy
14.6K
Super-resolution fluorescence microscopy (SRFM) provides a better resolution than conventional fluorescence microscopy by reducing the point spread function (PSF). PSF is the light intensity distribution from a point that causes it to appear blurred. Due to PSF, each fluorescing point appears bigger than its actual size, and it is the PSF interference of nearby fluorophores that causes the blurred image. Various approaches to achieving higher resolution through SRFM have recently been...