Topogram-based tube current modulation of head computed tomography for optimizing image quality while protecting the
Ming-Fang Lin1,2, Chia-Yuen Chen1,3, Yuan-Hao Lee1
11 Department of Radiology, Wan Fang Hospital, Taipei Medical University, Taipei, Taiwan.
CT eye shielding and topogram-based tube current modulation (TCM) significantly reduce radiation dose to the lens during head CT scans. This combination also improves image quality for nasal and periorbital imaging.
Area of Science:
- Radiology
- Medical Physics
- Radiation Protection
Background:
- Repeated head CT scans elevate the risk of radiation-induced lens opacification.
- Ocular radiation exposure is a concern in head and neck CT imaging.
Purpose of the Study:
- To evaluate the impact of CT eye shielding and topogram-based tube current modulation (TCM) on lens radiation dose.
- To assess the effect of these techniques on the image quality of nasal and periorbital regions.
Main Methods:
- An anthropomorphic phantom underwent CT scans with fixed or automatic tube current modulation.
- Lens radiation dose was measured using Gafchromic films with and without orbital shielding.
- Image quality was assessed via signal-to-noise ratio (SNR) and contrast-noise ratio (CNR), with artifact evaluation by CT specialists.
Main Results:
- Barium sulfate and bismuth-antimony shields with fixed tube current reduced lens radiation dose by 22.0% and 35.6%, respectively.
- Topogram-based TCM reduced artifacts from shields and improved SNR (21.6%) and CNR (7.2%).
- Combined TCM and shielding reduced lens doses by 12.2% (barium sulfate) and 27.2% (bismuth-antimony).
Conclusions:
- Topogram-based TCM effectively mitigates artifacts from bismuth-antimony shields used for lens radioprotection.
- This combination enhances SNR in periorbital tissues and nasal bone imaging.
- CT eye shielding and TCM offer a dual benefit of reduced lens dose and maintained/improved image quality.
More Related Videos
04:48Application of Deep Learning-Based Medical Image Segmentation via Orbital Computed Tomography
Published on: November 30, 2022
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Diamagnetic Shielding of Nuclei: Local Diamagnetic Current
Transmission-based Precautions II: Airborne and Protective Environment
Airborne precautions:
Use airborne precautions when treating patients known or suspected to have diseases that spread through the air—for example, tuberculosis or measles. These organisms are present in smaller droplets expelled by an infected person and...
Muscles that Move the Head
The bilateral sternocleidomastoid, or SCM, and the suprahyoid and infrahyoid muscles are significant head flexors. The SCM muscles originate at the sternum and clavicle and attach to the mastoid process of the temporal bone. The SCM contracts bilaterally to bend the head forward, whereas...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET