Non-Cartesian GRAPPA and coil combination using interleaved calibration data - application to concentric-ring MRSI of
Philipp Moser1, Wolfgang Bogner1, Lukas Hingerl1
1High Field MR Center, Department of Biomedical Imaging and Image-Guided Therapy, Medical University Vienna, Vienna, Austria.
Purpose:
Proton MR spectroscopic imaging (MRSI) benefits from B0 ≥ 7T and multichannel receive coils, promising substantial resolution improvements. However, MRSI acquisition with high spatial resolution requires efficient acceleration and coil combination. To speed up the already-fast sampling via concentric rings, we implemented additional, non-Cartesian, hybrid through-time/through-k-space (tt/tk)-generalized autocalibrating partially parallel acquisition (GRAPPA). A new multipurpose interleaved calibration scan (interleaved MUSICAL) acquires reference data for both coil combination and PI. This renders the reconstruction process (especially PI) less sensitive to instabilities.
Methods:
Six healthy volunteers were scanned at 7T. Three calibration datasets for coil combination and PI were recorded: a) iMUSICAL, b) static MUSICAL as prescan, c) moved MUSICAL as prescan with misaligned head position. The coil combination performance, including motion sensitivity, of iMUSICAL was compared to MUSICAL for single-slice free induction decay (FID)-MRSI. Through-time/through-k-space-GRAPPA with constant/variable-density undersampling was evaluated on the same data, comparing the three calibration datasets. Additionally, the proposed method was successfully applied to 3D whole-brain FID-MRSI.
Results:
Using iMUSICAL for coil combination yielded the highest signal-to-noise ratio (SNR) (+9%) and lowest Cramer-Rao lower bounds (CRLBs) (-6%) compared to both MUSICAL approaches, with similar metabolic map quality. Also, excellent mean g-factors of 1.07 and low residual lipid aliasing were obtained when using iMUSICAL as calibration data for two-fold, variable-density undersampling, while significantly degraded metabolic maps were obtained using the misaligned MUSICAL calibration data.
Conclusion:
Through-time/through-k-space-GRAPPA can accelerate already time-efficient non-Cartesian spatial-spectral 2D/3D-MRSI encoding even further. Particularly promising results have been achieved using iMUSICAL as a robust, interleaved multipurpose calibration for MRSI reconstruction, without extra calibration prescan.
More Related Videos
Related Concept Videos
Cartesian Vector Notation
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Cartesian Form for Vector Formulation
Vector Components in the Cartesian Coordinate System
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Plotting and Calibrating the Root Locus
The maximum gain occurs at the breakaway points between open-loop poles on the real axis, while the minimum gain is...