XEN Implant Fracture During Needling Procedure
Sara Olivari1,2, Carlo A Cutolo1,2, Letizia Negri1,2
1Eye Clinic, DiNOGMI, University of Genoa.
Xen implant fractures can occur during glaucoma surgery needling. Despite this complication, the Xen implant remains effective in lowering intraocular pressure.
Area of Science:
- Ophthalmology
- Surgical Devices
Background:
- The XEN implant is a hydrophilic stent used for glaucoma treatment.
- Postoperative needling is a common procedure for bleb-forming surgeries.
Observation:
- Three cases of XEN implant fracture occurred during the needling procedure in a series of 170 XEN implantations.
- The subconjunctival portion of the XEN implant fractured in these cases.
- The mean length of the fractured distal segment was 0.83 mm.
Findings:
- Despite XEN fracture in 3.1% of needling procedures, intraocular pressure (IOP) significantly decreased from a mean of 25.0 mmHg to 12.0 mmHg.
- No vision-threatening complications were observed during a mean follow-up of 15.3 months.
- Fracture of the XEN implant did not appear to compromise its efficacy in reducing IOP.
Implications:
- XEN implant fracture is a potential adverse event during bleb management that should be considered.
- The flexible material of the XEN implant makes it susceptible to needle damage.
- While fracture may not impair function, preserving implant integrity is recommended.
Related Concept Videos
06:59Fracture Apparatus Design and Protocol Optimization for Closed-stabilized Fractures in Rodents
08:18Three-Dimensional Reconstruction of Orbital Fractures
03:57Transverse Fracture of the Mouse Femur with Stabilizing Pin
Intraosseous Needle Placement
For unstable patients requiring urgent administration of medications, fluids, or blood products, establishing vascular access quickly is essential. However, there are many factors that can complicate placement of a peripheral intravenous cannula (PIV), and it is extremely common for PIV attempts to fail. PIV placement may be technically challenging in...
05:46Establishing a Diaphyseal Femur Fracture Model in Mice
03:56Operation Procedure and Precautions of Lingnan Fire-Needle Therapy in Osteoporosis Model Rats