Homogeneous Binary Visual and Fluorescence Detection of Tetanus Toxoid in Clinical Samples Based on Enzyme-Free
Piaopiao Chen1, Yunjin Bai1, Shiyuan Tang1,2
1Department of Laboratory Medicine, Med+X Center for Manufacturing, West China Precision Medicine Industrial Technology Institute, Department of Urology, Department of Emergency Medicine, Laboratory of Emergency Medicine, West China Hospital, Sichuan University, Chengdu, Sichuan 610041, China.
A new aptamer-based toxoid test offers sensitive detection of tetanus toxoid. This simple test uses fluorescence and visual readouts, validated in clinical samples for effective tetanus infection diagnosis.
Area of Science:
- Biomedical Engineering
- Nanotechnology
- Analytical Chemistry
Background:
- CdTe quantum dots can differentiate DNA-templated copper nanoparticles from Cu2+ ions.
- Aptamer-based assays offer high specificity for target biomolecules.
- Developing rapid and sensitive diagnostic tools for infectious diseases is crucial.
Purpose of the Study:
- To develop a simple aptamer-based toxoid test with dual readout capabilities.
- To achieve highly sensitive detection of tetanus toxoid.
- To validate the test in clinical settings and for patient monitoring.
Main Methods:
- Integration of CdTe quantum dots with aptamers for specific binding.
- Utilizing enzyme-free triple parallel hybridization chain reaction and cation exchange reaction.
- Employing inkjet printing for assay fabrication.
- Developing fluorescence and binary visual readouts (color and distance-based).
Main Results:
- Achieved a limit-of-detection (LOD) of 0.25 fg/mL for tetanus toxoid using fluorescence.
- Obtained LODs of 10 fg/mL (color) and 1 fg/mL (distance) for visual readouts.
- Successfully validated quantitative results in 84 clinical serum samples.
- Demonstrated accurate monitoring of tetanus toxoid levels in patients undergoing drug treatment.
Conclusions:
- The developed aptamer-based toxoid test provides a sensitive and versatile platform for tetanus toxoid detection.
- The dual readout system (fluorescence and visual) enhances applicability in diverse settings.
- A cutoff value of 5 fg/mL is recommended for diagnosing tetanus infection based on clinical validation.