[Comparative evaluation of 3 automated optical coagulation detection systems]
Annales De Biologie Clinique
|January 1, 1987
Summary
New automated coagulation analyzers offer reliable fibrinogen determination and high throughput. These advanced instruments provide excellent precision for Prothrombin Time and good results for Activated Partial Thromboplastin Time, showing equivalent performance to older devices.
Area of Science:
- Clinical Chemistry
- Hematology
- Medical Technology
Background:
- Traditional photo-optical coagulation analyzers have limitations in throughput and automation.
- The development of advanced automated hemostasis testing systems is crucial for efficient laboratory diagnostics.
Purpose of the Study:
- To evaluate and compare the performance of three new automated coagulation analyzers: Organon Teknika-General Diagnostics Coag A Mate X2 (CAM X2), Ortho Koagulab 40A, and MLA Electra 700 (E 700).
- To assess their reliability, accuracy, and efficiency in routine coagulation testing.
Main Methods:
- Simultaneous evaluation of the three analyzers using the same protocol and samples.
- Assessment of Prothrombin Time (PT), Activated Partial Thromboplastin Time (APTT), fibrinogen determination, and specific factor assays.
- Statistical analysis of repeatability, linearity, and correlation with existing methods.
Main Results:
- All three analyzers demonstrated high throughput and reliable fibrinogen determination with excellent linearity.
- Excellent repeatability was observed for PT and prothrombin complex factor assays (CV: 1-2.5%).
- Good repeatability was noted for APTT (CV: 0.5-3%) with specific reagent limitations, and fibrinogen assays yielded good results (CV: 3-4% and 6-8%).
- High correlation (R > 0.96) was found with results from other devices.
- The MLA Electra 700 (E 700) showed limitations in speed for APTT, while the CAM X2 was the quickest instrument.
Conclusions:
- The evaluated automated coagulation analyzers offer significant advantages in throughput, automation, and result expression compared to older devices.
- These instruments provide reliable and accurate results for key coagulation parameters, suitable for both routine and emergency testing.
- The MLA Electra 700 (E 700) is well-suited for emergencies due to its loading system, while the CAM X2 excels in handling large sample volumes.