Multi-analyte Biochip MAB Based on All-solid-state Ion-selective Electrodes ASSISE for Physiological Research
Published on: April 18, 2013
Dual-mode ion-selective electrodes and distance-based microfluidic device for detection of multiple urinary
Kamonchanok Phoonsawat1,2,3, Tugba Ozer4, Wijitar Dungchai1
1Organic Synthesis, Electrochemistry & Natural Product Research Unit, Department of Chemistry, Faculty of Science, King Mongkut's University of Technology Thonburi, Bangkok, 10140, Thailand.
Abstract:
Here, we developed a microfluidic paper device by combining ion-selective electrodes (ISE) and a distance-based paper device (dPAD) for simultaneous potentiometric and colorimetric detection of urine electrolytes including K+, Na+ and Cl-. The working and reference electrode zones were coated with polystyrene as a non-ionic polymer to improve hydrophobic properties on the paper surface for fabrication of K+-ISE and Na+-ISE. The layer of polymer coating was optimized to enhance the sensitivity of the ISEs. Under optimized conditions, the electrode surfaces were modified with carbon black to improve the electrochemical characteristics of the ISEs. The ISEs showed good performance with sensitivities of 54.14 ± 3.94 mV per decade and 55.08 ± 1.15 mV per decade for K+ and Na+ within the linear concentration range 0.100 mM-100 mM K+ and 5 mM-1 M Na+, respectively. The limits of detection (LOD) were 0.05 mM and 1.36 mM for K+ and Na+, respectively. The linear working range of Cl- was 0.50 to 50 mM and the LOD and limit of quantification (LOQ) were found to be 0.16 ± 0.05 mM (3SD) and 0.53 ± 0.05 mM (10SD), respectively. The dual-mode ISE-dPAD was validated in human urine and recoveries were obtained as 90-108%, 94-105%, and 90-96% for K+, Na+, and Cl-, respectively, showing successful application of the developed device in a complex matrix. The ISE-dPAD has advantages including low-cost ($ 0.33 per test), eco-friendly, portability, simple operation, the need of low sample volume (100 μL), and simultaneous analysis on a single device.
More Related Videos
09:18Measurement of Extracellular Ion Fluxes Using the Ion-selective Self-referencing Microelectrode Technique
Published on: May 3, 2015
13:42Dry Film Photoresist-based Electrochemical Microfluidic Biosensor Platform: Device Fabrication, On-chip Assay Preparation, and System Operation
Published on: September 19, 2017
Related Concept Videos
Potentiometry: Membrane Electrodes
Ion-Exchange Chromatography
Interfacial Electrochemical Methods: Overview
Electrodes: Overview
There are two main types of electrodes in electrochemical cells. The first type, known as the working or indicator electrode, has a potential that is sensitive to the analyte's concentration and reacts to changes in...
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
High-Performance Liquid Chromatography: Types of Detectors