Simple In-House Ultra-High Performance Capillary Column Manufacturing with the FlashPack Approach
Published on: December 4, 2021
General Two-Step Method for the Fabrication of Covalent-Organic Framework-Bound Open-Tubular Capillary Columns for
Wen-Chao Deng1,2,3, Hai-Long Qian1,2,3, Cheng Yang3
1State Key Laboratory of Food Science and Resources, Jiangnan University, Wuxi 214122, China.
Abstract:
Covalent organic frameworks (COFs) are promising as stationary phases for gas chromatography (GC). The successful anchoring of COFs to the inner walls of the capillary with good uniformity is an important prerequisite to ensure the excellent separation performance of columns. However, current methods for the fabrication of COF-based capillary columns cannot always meet this requirement when faced with different COFs, which hampers the further development of COF-based GC stationary phases. Here, we show a general two-step method for the fabrication of COF-bound capillary column. The first step enables the formation of uniform amorphous polymer layer on the inner walls of capillary, while the second step allows the facile transformation of the amorphous polymer layer into a highly crystalline COF layer. COF-bound capillary columns with different framework structures were fabricated successfully by the developed two-step method. Impressively, the COF layers bound on the inner walls of these capillary columns showed good uniformity and high crystallinity. More importantly, as an example, the fabricated Tab-DHTA-bound capillary column showed good resolution (R > 1.5) and high column efficiency (700-39,663 plates m-1) for the tested isomers of ethylbenzene, xylene, dichlorobenzene, chlorotoluene, pinene, 1,3-dichloropropene, and propylbenzene with good precision (RSD, run-to-run, n = 5) (retention time, 0.2-0.6%; peak area, 0.5-1.1%; and peak height, 0.5-1.4%). In general, the fabricated Tab-DHTA-bound capillary column exhibited better performance for the separation isomers than commercial columns DB-5 and HP-FFAP. These results indicate that the two-step method is an efficient way to fabricate the COF-bound capillary column with excellent separation performance.
More Related Videos
11:44Qualitative Characterization of the Aqueous Fraction from Hydrothermal Liquefaction of Algae Using 2D Gas Chromatography with Time-of-flight Mass Spectrometry
Published on: March 6, 2016
08:15Metabolomic Analysis of Barley by Gas Chromatography/Mass Spectrometry
Published on: November 8, 2024
Related Concept Videos
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Principles Of Column Chromatography
High-Performance Liquid Chromatography: Elution Process
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...