The Effect of Laser-Resistant Endotracheal Tube Design on Airflow Dynamics: A Benchtop and Clinical Study
Jessica M L Pagel1, Adithya Reddy1, Lucy Fitzgerald2
1Department of Otolaryngology-Head and Neck Surgery, University of Virginia, Charlottesville, VA, USA.
Objective:
Compare ventilation pressures of 2 endotracheal tube designs used in laser airway surgery in clinical practice and with a benchtop model to elucidate differences and understand the design elements that impact airflow dynamics.
Methods:
Ventilatory and aerodynamic characteristics of the laser resistant stainless-steel endotracheal tube (LRSS-ET) design and the laser resistant aluminum-wrapped silicone endotracheal tube (LRAS-ET) design were compared. Ventilatory parameters were collected for 32 patients undergoing laser-assisted airway surgery through retrospective chart review. An in vitro benchtop simulation measured average resistance and centerline turbulence intensity of both designs at various diameters and physiological frequencies.
Results:
Baseline patient characteristics did not differ between the 2 groups. Clinically, the median LRAS-ET peak inspiratory pressure (PIP; 21.00 cm H2O) was significantly decreased compared to LRSS-ET PIP (34.67 cm H2O). In benchtop simulation, the average PIP of the LRAS-ET was significantly lower at all sizes and frequencies. The LRSS-ET consistently demonstrated an increased resistance, although no patterns were observed in turbulence intensity data between both designs.
Conclusion:
The benchtop model demonstrated increased resistance in the LRSS-ET compared to the LRAS-ET at all comparable sizes. This finding is supported by retrospective ventilatory pressures during laser airway surgery, which show significantly increased PIPs when comparing identically sized inner diameters. Given the equivocal turbulence intensity data, these differences in resistance and pressures are likely caused by wall roughness and intraluminal presence of tubing, not inlet or outlet geometries. The decreased PIPs of the LRAS-ET should assist in following lung protective ventilator management strategies and reduce risk of pulmonary injury and hemodynamic instability to the patient.
Related Concept Videos
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Endotracheal Intubation II: Nursing Management
1. Nursing Care of Patients Before Intubation
Before the endotracheal intubation procedure, nurses play an essential role in ensuring the process goes smoothly. The nurses must be familiar with intubation...