Computed tomography numbers obtained for varying iodine contrast concentrations by different-generation dual-energy
Tomokatsu Tsukamoto1, Takanori Masuda2, Takashi Takahata1
1Department of Radiology, Onomichi General Hospital, 1-10-23 Hirahara, Onomichi City, Hiroshima Pref 722-8508, Japan.
Abstract:
We compared the computed tomography (CT) numbers from monochromatic images obtained using the first-generation (Discovery CT750 HD: GE Healthcare, Milwaukee, WI) and second-generation (Revolution CT: GE HealthCare) dual-energy CT (first and second DECT) scanners in phantom and clinical studies. In a polypropylene phantom, eight polypropylene tubes containing iodine at various concentrations (0.5, 1, 2, 5, 10, 12, 20, 30 mg I per ml) were arranged in an outer circle. The iodine densities and CT numbers obtained after imaging with different-generation DECT scanners were analyzed. The CT numbers from images obtained from 61 consecutive patients with aortic disease who underwent CT with different-generation DECT scanners were compared during the arterial and delayed phases. The iodine concentration obtained from second DECT was more accurate than that from the first DECT in the phantom study. A significantly higher contrast enhancement was observed with the second DECT compared with the first DECT during the arterial phase in the clinical study. Contrast enhancement was higher with the second DECT than with the first DECT, and the second DECT was effective in minimizing the use of contrast materials.
More Related Videos
08:41Novel In Vivo Micro-Computed Tomography Imaging Techniques for Assessing the Progression of Non-Alcoholic Fatty Liver Disease
Published on: March 24, 2023
08:13In Vivo Quantitative Assessment of Myocardial Structure, Function, Perfusion and Viability Using Cardiac Micro-computed Tomography
Published on: February 16, 2016
Related Concept Videos
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
X-ray Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...