A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Defibrillation Testing in Patients Undergoing Replacement of the S-ICD Generator: Is There Still a Need?
Benedikt Biller1, Florian Reinke1, Katharina Biller2
1Department of Cardiology II - Electrophysiology, University Hospital Münster, Münster, Germany.
Omitting defibrillation threshold (DFT) testing during subcutaneous-ICD (S-ICD) generator replacement is not recommended. The PRAETORIAN score and high-voltage impedance can predict conversion failure, guiding decisions on whether to perform DFT testing.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Device Technology
Background:
- Intraoperative defibrillation threshold (DFT) testing can be omitted for transvenous implantable cardioverter defibrillators (ICDs).
- Data regarding the necessity of DFT testing for subcutaneous-ICDs (S-ICDs), particularly during generator replacement, is lacking.
Purpose of the Study:
- To evaluate the outcome of intraoperative DFT during S-ICD generator replacement.
- To identify predictors of conversion failure during S-ICD generator replacement procedures.
Main Methods:
- Retrospective analysis of 112 consecutive patients undergoing S-ICD generator replacement and routine testing.
- Assessment of intraoperative DFT outcomes and correlation with patient characteristics, PRAETORIAN score, and high-voltage (HV) impedance.
Main Results:
- Successful DFT testing was achieved in 87.3% of patients during generator replacement without major adverse events.
- The PRAETORIAN score was generally lower, and HV impedance increased compared to initial implantation.
- Conversion failure risk increased with higher PRAETORIAN risk class; PRAETORIAN score and HV impedance were the most accurate predictors of failure.
Conclusions:
- Routine omission of DFT testing during S-ICD generator replacement is not advisable.
- New algorithms utilizing low impedance DFT measurements and the PRAETORIAN score can predict conversion failure.
- Patients with HV impedance <70 Ohms and PRAETORIAN score <90 have a very low risk of conversion failure. DFT testing is recommended for patients with PRAETORIAN score >150 during generator replacement.