Omnidirectional Wireless Power Transfer for Millimetric Magnetoelectric Biomedical Implants
Wei Wang1, Zhanghao Yu1, Yiwei Zou1
1Department of Electrical and Computer Engineering, Rice University, Houston TX, 77005, USA.
Summary
This study introduces an omnidirectional wireless power transfer (WPT) system for tiny bioelectronic implants. It overcomes misalignment issues, significantly improving power transfer efficiency for medical devices.
Area of Science:
- Bioelectronic Engineering
- Medical Device Technology
- Wireless Power Transfer
Background:
- Miniature bioelectronic implants offer advanced therapies for cardiovascular and neurological conditions.
- Wireless power transfer (WPT) is crucial for battery-free implants, but efficiency drops with misalignment.
- Existing WPT systems struggle with body movements and implant orientation, limiting reliability.
Purpose of the Study:
- To develop an omnidirectional WPT platform for millimetric bioelectronic implants.
- To enhance the robustness and efficiency of WPT despite misalignment.
- To enable reliable power delivery for implantable medical devices.
Main Methods:
- Utilized magnetoelectric (ME) WPT modality with magnetic field steering via multiple transmitter (TX) coils.
- Developed an Active Echo (AE) scheme with a miniature implant coil for closed-loop control of the TX array.
- Integrated a 14.2mm³ implantable stimulator with a custom SoC and an ME film, alongside a multi-coil TX array.
Main Results:
- The Active Echo (AE) receiver achieved -161dBm/Hz input-referred noise and 64dB gain tuning.
- The omnidirectional WPT system demonstrated 6.8x higher power transfer efficiency (PTE) at 90° misalignment compared to a single-coil system.
- AE scheme minimally impacted PTE (<2%) even with tracking errors.
Conclusions:
- The proposed omnidirectional ME WPT system with AE significantly improves power transfer robustness and efficiency for millimetric bioelectronic implants.
- This technology addresses key challenges in powering implantable devices, paving the way for more reliable therapies.
- The Active Echo scheme offers a robust solution for adaptive control in WPT systems, enhancing medical implant functionality.