Related Experiment Video
Updated: May 12, 2025

28:13
Catheter Ablation in Combination With Left Atrial Appendage Closure for Atrial Fibrillation
Published on: February 26, 2013
33.2K
Char Formation During Very High-Power Short-Duration Ablation for Atrial Fibrillation.
Kazumasa Suga1, Hiroyuki Kato1, Koudai Yamazaki1
1Department of Cardiology, Japan Community Healthcare Organization Chukyo Hospital, Nagoya, Japan.
Journal of Cardiovascular Electrophysiology
|May 9, 2025
Summary
Very high-power short-duration (vHPSD) ablation for atrial fibrillation can cause char formation in nearly half of patients. Avoiding specific anatomical sites and catheter orientations may reduce this risk.
Area of Science:
- Cardiology
- Electrophysiology
- Medical Devices
Background:
- Pulmonary vein (PV) isolation using very high-power short-duration (vHPSD) radiofrequency catheter ablation is a key treatment for atrial fibrillation.
- vHPSD ablation carries a risk of char formation, necessitating investigation into its incidence and predictors.
Purpose of the Study:
- To assess the incidence of char formation during vHPSD pulmonary vein isolation (PVI).
- To identify patient characteristics and procedural factors associated with char formation during vHPSD-PVI.
Main Methods:
- Prospective study of 50 patients undergoing initial PVI with a QDOT MICRO™ catheter in vHPSD setting (90W/4s).
- PVs were divided into eight segments; char formation was evaluated post-ablation.
- Analysis of patient characteristics and procedural parameters to identify risk factors.
Main Results:
- Char formation occurred in 48.0% of patients, predominantly in the right inferior PV segments.
- Higher left atrial volume index (LAVI) and parallel catheter orientation were independent predictors of char formation.
- One patient with char formation experienced symptomatic thromboembolism.
Conclusions:
- Char formation is a significant complication of vHPSD-PVI, observed in 48.0% of patients.
- Risk factors include application to the right inferior PV, higher LAVI, and parallel catheter orientation.
- vHPSD applications should be approached cautiously in these specific conditions to mitigate char formation risk.