Capillary Blood Self-Sampling for Therapeutic Drug Monitoring: A Mixed-Methods Usability Study of the True Dose® Kit
Marie-Therése Crafoord1, Elham Hedayati2,3, Aina Johnsson2
1Department of Clinical Science and Education, Södersjukhuset (South Hospital), Karolinska Institutet, Stockholm, Sweden.
Biomed Research International
|July 18, 2025
Summary
The True Dose kit shows high usability for capillary blood self-sampling, though instructions and stress management need improvement for better user experience and consistent performance.
Area of Science:
- Medical Device Usability Engineering
- Human Factors in Healthcare
- Diagnostic Self-Sampling Technologies
Background:
- Usability is critical for medical device safety and effectiveness, particularly for self-sampling kits.
- The True Dose kit aims to facilitate capillary blood self-sampling, requiring rigorous usability evaluation.
- Adherence to EN 62366 standards is essential for medical device usability validation.
Purpose of the Study:
- To evaluate the usability of the True Dose kit for capillary blood self-sampling according to EN 62366.
- To assess user performance, experience, and confidence with the True Dose kit.
- To identify areas for improvement in the True Dose kit's design and instructions.
Main Methods:
- A convergent parallel mixed-methods design integrating quantitative and qualitative data.
- Twenty-five participants performed capillary blood self-sampling using the True Dose kit.
- Quantitative metrics (blood volume, air bubbles, duration) and qualitative data (interviews, feedback) were collected, alongside System Usability Scale (SUS) and Test Feedback Questionnaire (TFQ) scores.
Main Results:
- The True Dose kit achieved a high overall SUS score of 82.70, indicating good usability.
- 92% of participants successfully collected adequate blood samples, with a median sampling time of 10.42 minutes.
- Participants reported challenges with instruction clarity and managing stress during blood flow, suggesting a need for enhanced visual guidance and feedback.
Conclusions:
- The True Dose kit demonstrates significant usability potential for capillary blood self-sampling.
- Improvements in instructional design and stress management strategies are recommended to optimize user experience.
- Further refinement can enhance consistent performance and user confidence in self-sampling procedures.
Keywords:
blood chemical analysisblood specimen collectionclinical trialsdrug approvalinstructional film and videouser-centered design