Related Experiment Video
Updated: Jul 21, 2026

In situ Compressive Loading and Correlative Noninvasive Imaging of the Bone-periodontal Ligament-tooth Fibrous Joint
Published on: March 7, 2014
Investigation of the minimum uptake time for quantitative jawbone SPECT.
Hironobu Hata1, Kenji Hirata2, Satoshi Shimomura3
1Department of Dentistry and Oral Surgery,, NHO Hokkaido Cancer Center, Sapporo, Japan. hata.hironobu.wq@mail.hosp.go.jp.
Reducing uptake time for jawbone SPECT using [99mTc] Tc-HMDP is beneficial for cancer patients. However, this study found that the shortest acceptable acquisition time for quantitative jawbone SPECT remains 2 hours.
Area of Science:
- Nuclear Medicine
- Radiopharmacology
- Medical Imaging
Background:
- Bone scintigraphy imaging times vary, impacting patient care.
- Previous studies showed [99mTc] Tc-HMDP uptake at 2 hours is comparable to 3 hours.
- Shorter uptake times benefit patients with poor general health, including cancer patients.
Purpose of the Study:
- To investigate if uptake time for quantitative jawbone SPECT using [99mTc] Tc-HMDP can be further reduced.
- To determine the minimum acceptable uptake time for accurate quantitative SPECT analysis of jawbone lesions.
Main Methods:
- Forty patients underwent jawbone SPECT imaging with [99mTc] Tc-HMDP.
- Patients were divided into 1.0-hour and 1.5-hour uptake time groups, with imaging at 1.0/2.0h and 1.5/2.0h respectively.
- Maximum standardized uptake values (SUVmax) were compared between time points and groups using quantitative SPECT software.
Main Results:
- In the 1.0-hour group, parietal bone SUVmax showed no significant difference, but jawbone lesion SUVmax significantly increased by 2.0h (p < 0.001).
- In the 1.5-hour group, parietal bone SUVmax was not significantly different, while jawbone lesion SUVmax significantly increased by 2.0h (p < 0.001).
- No significant difference was found in parietal bone uptake between earlier and later time points in either group.
Conclusions:
- Quantitative SPECT evaluation of jawbone lesions using [99mTc] Tc-HMDP cannot be reduced by 30-minute increments from 2 hours.
- The shortest acceptable acquisition time for quantitative jawbone SPECT remains 2 hours.
- Further reduction in uptake time below 2 hours is not feasible for accurate quantitative analysis.