In Vitro Evaluation of Apical Transportation Induced by Single- and Dual-Thermally Treated NiTi Instruments
Francesco Puleio1, Rosario Pirri2, Jose Aranguren3
1Dipartimento di Scienze Biomediche Odontoiatriche e delle Immagini Morfologiche e Funzionali, Università degli Studi di Messina, Messina, SIcily, Italy, unime.it.
Dual-thermally treated SlimShaper PRO nickel-titanium (NiTi) endodontic instruments did not increase apical transportation compared to single-thermally treated SlimShaper instruments. This study evaluated apical transportation in simulated canals using these NiTi instruments.
Area of Science:
- Endodontics
- Dental Materials Science
- Biomaterials Engineering
Background:
- Martensitic endodontic instruments offer superior flexibility, reducing apical transportation versus austenitic instruments.
- Newer instruments with dual thermal treatment aim for enhanced tip cutting and body flexibility.
- Concerns exist that increased tip stiffness may elevate apical transportation.
Purpose of the Study:
- To compare apical transportation between SlimShaper (single thermal treatment) and SlimShaper PRO (dual thermal treatment) nickel-titanium (NiTi) instruments.
- To assess the impact of dual thermal treatment on NiTi endodontic instrument performance regarding apical transportation.
Main Methods:
- Forty-four standardized J-shaped resin blocks were used, divided into two groups (n=22).
- Group A utilized SlimShaper, and Group B used SlimShaper PRO, both up to size 25/0.04.
- Apical transportation was evaluated by superimposing pre- and post-shaping photographs; statistical analysis employed the independent Student's t-test (p < 0.05).
Main Results:
- Mean apical transportation for SlimShaper was 0.08423 mm (SD 0.01049).
- Mean apical transportation for SlimShaper PRO was 0.08609 mm (SD 0.01102).
- No statistically significant difference in apical transportation was found between the two groups (p = 0.57).
Conclusions:
- Dual-thermally treated SlimShaper PRO instruments did not result in greater apical transportation than single-thermally treated SlimShaper instruments.
- The findings suggest that dual thermal treatment, under the study conditions, does not adversely affect apical transportation in NiTi endodontic instruments.
Related Concept Videos
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for:
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
Instrument Transformers
Uncertainty in Measurement: Reading Instruments
Facilitated Transport