Intraocular pressure in Lewis rats
A Mermoud1, G Baerveldt, D S Minckler
1Doheny Eye Institute, Los Angeles, CA 90033.
Investigative Ophthalmology & Visual Science
|April 1, 1994
Summary
Accurate noninvasive intraocular pressure (IOP) measurements in rats are possible using Tono-Pen tonometers. This study established normal IOP values and confirmed no significant long-term fluctuations in Lewis rats.
Area of Science:
- Ophthalmology
- Veterinary Medicine
- Physiology
Background:
- Intraocular pressure (IOP) is a critical factor in ocular health.
- Accurate IOP measurement in animal models is essential for research.
- Noninvasive methods are preferred for repeated measurements in research settings.
Purpose of the Study:
- To calibrate Tono-Pen-1 and Tono-Pen-2 tonometers against direct manometry for noninvasive IOP measurement in rats.
- To establish normal IOP values in Lewis rats.
- To determine long-term IOP fluctuations in Lewis rats.
Main Methods:
- Calibration involved cannulating rat eyes and connecting them to a pressure transducer, with IOP incrementally increased and measured by tonometers.
- Normal IOP was determined by measuring IOP in 229 eyes of 115 Lewis rats using Tono-Pen-1.
- Long-term fluctuations were assessed by daily IOP measurements over 7 days in 52 eyes of 26 rats.
Main Results:
- Regression formulas were derived for Tono-Pen-1 (y = 1.819 + 0.711x, r2 = 0.92) and Tono-Pen-2 (y = -1.291 + 0.784x, r2 = 0.97).
- Normal IOP in rats was determined to be 17.30 +/- 5.25 mm Hg, with a 90% confidence interval of 7.28 to 26.98 mm Hg.
- No significant long-term fluctuations in IOP were observed (P = 0.55).
Conclusions:
- Tono-Pen-1 and Tono-Pen-2 tonometers provide accurate noninvasive measurements of IOP in living rats.
- The established normal IOP range and lack of fluctuation aid in future rat ocular research.