Physical and subjective evaluation of a three-detector (TRIAD 88) SPECT system
M F D'Souza1, C G Mumma, S M Mohapatra
1Department of Radiological Sciences, University of Oklahoma Health Sciences Center, Oklahoma City 73190, USA.
Australasian Physical & Engineering Sciences in Medicine
|December 1, 1995
Summary
The TRIAD 88 SPECT system
Area of Science:
- Nuclear Medicine
- Medical Imaging Physics
Background:
- The TRIAD 88 is a whole-body SPECT system with variable cylindrical FOV.
- It is designed for both brain and body organ imaging.
Purpose of the Study:
- To assess the physical and clinical performance of the TRIAD 88 SPECT system.
- To compare two parallel hole collimator types: LEUR_PAR and LEHR_PAR.
Main Methods:
- Evaluated physical indices: contrast-detail curves, LSFs, sensitivity, energy resolution, uniformity.
- Used Carlson phantom, Hoffman brain phantom, and clinical brain images for subjective comparison.
- Analyzed display characteristics using H & D curves.
Main Results:
- LEUR_PAR collimator had 58% of the sensitivity of LEHR_PAR.
- Planar spatial resolution differed by 14% between collimators.
- No statistical difference in lesion detectability on planar contrast-detail curves, but improved performance with LEUR_PAR on phantom and clinical SPECT images.
- Consistent performance across the three detector heads.
Conclusions:
- The TRIAD 88 system demonstrates good performance for brain and body imaging.
- While sensitivity differs, the LEUR_PAR collimator shows improved image quality in phantom and clinical SPECT imaging.
- The system exhibits consistent performance among its three detector heads.
More Related Videos
Related Concept Videos
UV–Vis Spectrometers
The absorbance of UV and visible (UV–visible) radiations is measured using a UV–visible spectrophotometer. Deuterium lamps, which emit UV radiation, and tungsten lamps, which produce radiation in the visible region, are used as light sources in UV–visible spectrophotometers. A monochromator or prism is used for diffraction grating, i.e., to split the incoming radiation into different wavelengths. A system of slits is used to focus the desired wavelength on the sample cell. Samples for...
IR Spectrometers
There are two main infrared (IR) spectrophotometers: dispersive IR spectrometers and Fourier transform infrared (FTIR) spectrometers. In a dispersive IR spectrometer, a beam of infrared radiation produced by a hot wire is divided into two parallel equal-intensity beams using mirrors. One beam passes through the sample, while another is a reference beam. The beams then move through the monochromator, which separates the radiations into a continuous spectrum of different frequencies. The...
Mass Analyzers: Common Types
The quadrupole mass analyzer consists of four cylindrical metal rods arranged in a diamond carrying a DC voltage and a radio-frequency AC voltage. The motion of ions through the quadrupole depends on the field strength, causing only ions of a certain m/z to resonate successfully and strike the detector at a given field strength. Though the transmission rate for these analyzers is high, the exact elemental composition of the sample is not determined because of low resolution; however, they are...
Gas Chromatography: Types of Detectors-II
In gas chromatography, different detectors are employed to meet specific analytical needs. These detectors are often categorized based on their detection mechanisms and the types of compounds they are best suited to analyze. Thermal Conductivity Detectors (TCD), Flame Ionization Detectors (FID), and Electron Capture Detectors (ECD) represent common categories, each with unique operating principles and applications. However, beyond these, several other detectors are designed for more specialized...
High-Performance Liquid Chromatography: Types of Detectors
The role of the detectors in High-Performance Liquid Chromatography (HPLC) is to analyze the solutes as they exit from the chromatographic column. The detector recognizes the solute's property and generates corresponding electrical signals, which are converted into a readable graph of the detector's response versus elution time called a chromatogram at the computer. There are several types of HPLC detectors, each with its own advantages and limitations, depending on the analyte properties and...
Atomic Absorption Spectroscopy: Instrumentation
An atomic absorption spectrophotometer (AAS) comprises several components: a radiation source, an atomizer, a monochromator, and a detector. The radiation source can be a hollow-cathode lamp (HCL) or an electrodeless-discharge lamp (EDL), both of which provide a narrow emission line of the required wavelength. However, some instruments use continuum sources and high-resolution monochromators to achieve a narrow range of radiation.
The atomizer used in AAS can be either a flame atomizer or an...
The atomizer used in AAS can be either a flame atomizer or an...