Copper T 380A IUD and magnetic resonance imaging
1Department of Gynecology and Obstetrics, Emory University School of Medicine, Atlanta, GA 30303, USA. mzieman@comcast.net
The Copper T 380A intrauterine device (IUD) is safe for 3.0 Tesla (T) magnetic resonance (MR) imaging. Testing showed no adverse effects, confirming its "MR Conditional" status for this higher field strength.
Area of Science:
- Medical Device Safety
- Magnetic Resonance Imaging
- Reproductive Health
Background:
- Women with implanted medical devices require safety data for MRI procedures.
- Previous testing of the Copper T 380A IUD was limited to 1.5 T MR systems.
- Higher field strength MRI necessitates updated safety evaluations for IUDs.
Purpose of the Study:
- To evaluate the safety of the Copper T 380A IUD at 3.0 T MR imaging.
- To determine potential risks associated with 3.0 T MR exposure for this IUD.
- To assess compatibility with current medical device safety standards.
Main Methods:
- In vitro safety testing of the Copper T 380A IUD.
- Utilized a GE Signa LX 3.0-T MR system for testing.
- Evaluated variables including deflection, torque, heating, and artifact generation.
Main Results:
- No significant deflection or torque was observed.
- No detrimental heating effects were detected.
- No significant artifact generation was found in the testing model.
Conclusions:
- The Copper T 380A IUD demonstrated no safety concerns at 3.0 T under tested conditions.
- Results align with new ASTM and FDA standards for MR Conditional devices.
- The IUD is considered safe for 3.0 T MR imaging based on these findings.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation I: X-ray and CT
Intrauterine Drug Delivery Systems