A Simple Approach to Perform TEER Measurements Using a Self-Made Volt-Amperemeter with Programmable Output Frequency
Published on: October 5, 2019
A novel in vivo corneal trans-epithelial electrical resistance measurement device
Masafumi Uematsu1, Yasser Helmy Mohamed2, Naoko Onizuka1
1Department of Ophthalmology and Visual Sciences, Graduate School of Biomedical Sciences, Nagasaki University, 1-7-1 Sakamoto, Nagasaki, Nagasaki 852-8501, Japan.
Purpose:
To develop a device that is capable of easily measuring corneal transepithelial electrical resistance (TER) and changes in the corneal barrier function.
Methods:
We had previously developed an in vivo method for measuring corneal TER using intraocular electrode. This method can be used to precisely measure the decline of the corneal barrier function after instillation of benzalkonium chloride (BAC). In order to lessen the invasiveness of that procedure, we further refined the method for measuring the corneal TER by developing electrodes that could be placed on the cornea and in the conjunctival sac instead of inserting them into the anterior chamber. TER was then calculated by subtracting the electrical resistance, which lacked the corneal epithelial input, from the whole electrical resistance that was measured between the electrodes. Slit lamp examination and scanning electron microscopy (SEM) were used to determine safety of the new device. Corneal TER changes after exposure to 0.02% BAC were determined using the new device as well as SEM and transmission electron microscopy (TEM).
Results:
Slit lamp examination before and after exposure of rabbits' corneas to the sensor confirmed safety of the device. SEM examination revealed no difference of the corneal epithelium which exposed to the new device with normal corneas. SEM and TEM pictures revealed damaged microvilli and tight junctions after instillation of 0.02% BAC. TER change after treatment with 0.02%BAC was similar to those determined by the established anterior chamber method.
Conclusion:
We succeeded to develop a less invasive device for corneal TER measurement in vivo in animals. This new device may be applicable in the future for clinical use in humans.