Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences01:17

NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences

1.5K
A pulse is a short burst of radio waves distributed over a range of frequencies that simultaneously excites all the nuclei in the sample. Upon passing a radio frequency pulse along the x-axis, the nuclei absorb energy corresponding to their Larmor frequencies and achieve resonance. This shifts the net magnetization vector from the z-axis toward the transverse plane. This angle of rotation of the magnetization vector, or the flip angle, is proportional to the duration and intensity of the pulse.
1.5K
Magnetic Resonance Imaging01:24

Magnetic Resonance Imaging

8.8K
Magnetic resonance imaging (MRI) is a noninvasive medical imaging technique based on a phenomenon of nuclear physics discovered in the 1930s, in which matter exposed to magnetic fields and radio waves was found to emit radio signals. In 1970, a physician and researcher named Raymond Damadian noticed that malignant (cancerous) tissue gave off different signals than normal body tissue. He applied for a patent for the first MRI scanning device in clinical use by the early 1980s. The early MRI...
8.8K
Assessment of Diffusion and Perfusion01:17

Assessment of Diffusion and Perfusion

1.4K
Understanding and evaluating diffusion and perfusion is critical in assessing a patient's respiratory and circulatory health. These processes play key roles in maintaining the body's internal environment, ensuring that tissues receive adequate oxygen while waste products are efficiently removed.
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
1.4K
Double Resonance Techniques: Overview01:12

Double Resonance Techniques: Overview

581
Double resonance techniques in Nuclear Magnetic Resonance (NMR) spectroscopy involve the simultaneous application of two different frequencies or radiofrequency pulses to manipulate and observe two distinct nuclear spins. One important application of double resonance is spin decoupling, which selectively suppresses coupling with one type of nucleus while observing the NMR signal from another nucleus, simplifying the spectrum and enhancing resolution.
Spin decoupling is usually achieved by...
581
Diffusion01:12

Diffusion

215.1K
Diffusion is the passive movement of substances down their concentration gradients—requiring no expenditure of cellular energy. Substances, such as molecules or ions, diffuse from an area of high concentration to an area of low concentration in the cytosol or across membranes. Eventually, the concentration will even out, with the substance moving randomly but causing no net change in concentration. Such a state is called dynamic equilibrium, which is essential for maintaining overall...
215.1K

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Characterizing Metabolic and Compositional Heterogeneity of Calf Muscle Using CEST MRI at 3 T.

NMR in biomedicine·2026
Same author

Metabolome-based genome-wide association study reveals PANK2 and EEF1A2 as major genetic regulators of key volatile compounds in Gaoyou duck breast meat.

Poultry science·2026
Same author

Age-Specific Contrast Optimization of bSSFP in Fetal Brain.

Magnetic resonance in medicine·2026
Same author

Microstructure imaging of prostate cancer by diffusion MRI.

Magma (New York, N.Y.)·2026
Same author

Molecular-Structural MRI Mismatch for Preoperative Risk Stratification of Pediatric Neuroblastoma.

Radiology. Imaging cancer·2026
Same author

SNR-Enhanced Whole-Brain Chemical Exchange Saturation Transfer Imaging by Optimized Variable Flip Angles.

Magnetic resonance in medicine·2026
Same journal

Feasibility and SNR Performance of Hyperpolarized <sup>129</sup>Xe Gas Exchange Imaging Using a Balanced SSFP Sequence.

Magnetic resonance in medicine·2026
Same journal

Multi-Contrast Human Brain CEST MRI at 11.7 T: First In Vivo Demonstration.

Magnetic resonance in medicine·2026
Same journal

Suppression of Oscillation and Ghosting in RF-Spoiled Gradient-Echo-Based Dynamic Imaging.

Magnetic resonance in medicine·2026
Same journal

A Simple, Dynamic Geometric Phantom for MRI and CT Reconstruction Pipelines: Beyond Shepp-Logan.

Magnetic resonance in medicine·2026
Same journal

7T 3D-EPI PCASL With High SNR Efficiency and Robustness to Through-Plane B<sub>0</sub> Field Gradients.

Magnetic resonance in medicine·2026
Same journal

A Comparison of Tissue Property Values Estimated Using Conventional Cardiac MRF and MT-Cardiac MRF.

Magnetic resonance in medicine·2026
See all related articles

Related Experiment Video

Updated: Dec 15, 2025

Registered Bioimaging of Nanomaterials for Diagnostic and Therapeutic Monitoring
17:16

Registered Bioimaging of Nanomaterials for Diagnostic and Therapeutic Monitoring

Published on: December 9, 2010

10.6K

Diffusion-prepared 3D gradient spin-echo sequence for improved oscillating gradient diffusion MRI.

Dan Wu1, Dapeng Liu2,3, Yi-Cheng Hsu4

  • 1Key Laboratory for Biomedical Engineering of Ministry of Education, College of Biomedical Engineering & Instrument Science, Zhejiang University, Hangzhou, Zhejiang, China.

Magnetic Resonance in Medicine
|July 10, 2020
PubMed
Summary
This summary is machine-generated.

This study introduces a new 3D imaging sequence that improves the quality and speed of specialized brain scans. By combining advanced signal preparation with efficient data collection, the method produces clearer images with less distortion on standard hospital scanners. This advancement helps doctors better measure how water moves in brain tissue, potentially improving clinical diagnostics.

Keywords:
3D-GRASEdiffusion-time dependencyoscillating gradientsignal-to-noise ratio | scan timeneuroimagingsignal-to-noise ratioGRASE sequencemicrostructure mapping

Frequently Asked Questions

More Related Videos

Metabolic Support of Excised, Living Brain Tissues During Magnetic Resonance Microscopy Acquisition
10:21

Metabolic Support of Excised, Living Brain Tissues During Magnetic Resonance Microscopy Acquisition

Published on: October 18, 2017

6.2K
Quantitative Magnetic Resonance Imaging of Skeletal Muscle Disease
09:30

Quantitative Magnetic Resonance Imaging of Skeletal Muscle Disease

Published on: December 18, 2016

20.0K

Related Experiment Videos

Last Updated: Dec 15, 2025

Registered Bioimaging of Nanomaterials for Diagnostic and Therapeutic Monitoring
17:16

Registered Bioimaging of Nanomaterials for Diagnostic and Therapeutic Monitoring

Published on: December 9, 2010

10.6K
Metabolic Support of Excised, Living Brain Tissues During Magnetic Resonance Microscopy Acquisition
10:21

Metabolic Support of Excised, Living Brain Tissues During Magnetic Resonance Microscopy Acquisition

Published on: October 18, 2017

6.2K
Quantitative Magnetic Resonance Imaging of Skeletal Muscle Disease
09:30

Quantitative Magnetic Resonance Imaging of Skeletal Muscle Disease

Published on: December 18, 2016

20.0K

Area of Science:

  • Medical imaging physics within diagnostic radiology
  • Advanced oscillating gradient diffusion MRI signal processing

Background:

Clinical adoption of time-dependent diffusion imaging remains limited by significant technical hurdles. Oscillating gradient methods allow researchers to probe short diffusion timescales within biological tissues. However, these sequences often suffer from poor signal quality and lengthy scan durations. That uncertainty drove the development of more efficient acquisition strategies for standard hospital hardware. Prior research has shown that conventional two-dimensional approaches struggle with image artifacts and low efficiency. No prior work had resolved the trade-off between scan speed and image clarity for these specific protocols. This gap motivated the creation of a three-dimensional preparation module for clinical use. The current investigation addresses these limitations by integrating a novel readout technique into the imaging pipeline.

Purpose Of The Study:

The aim of this study is to develop a 3D oscillating gradient-prepared gradient spin-echo sequence to enhance clinical diffusion imaging. Researchers sought to address the technical challenges associated with oscillating gradient methods on standard scanners. The primary motivation was to improve signal-to-noise ratio while simultaneously reducing total acquisition time. Current clinical protocols often struggle with lengthy scan durations and significant image artifacts. This work specifically targets the limitations of conventional 2D-EPI sequences in thick slice coverage applications. The team intended to demonstrate that a 3D approach could provide superior image quality for time-dependent diffusion measurements. By optimizing the readout and preparation modules, the authors aimed to facilitate broader clinical adoption. This investigation provides a systematic evaluation of the proposed sequence against established standards.

Main Methods:

Review approach involved comparing the performance of the novel 3D sequence against standard 2D-EPI protocols. Investigators implemented a sequence featuring global saturation, diffusion encoding, fat saturation, and GRASE readout modules. They utilized multiplexed sensitivity-encoding to rectify phase inconsistencies across multiple shots. The team evaluated scan efficiency and signal strength across 30-90-mm slice coverage. Researchers performed measurements at 50 Hz and 25 Hz for oscillating gradient protocols. They also conducted pulsed-gradient experiments at 30 ms and 60 ms diffusion times. Data collection occurred on a 3T clinical scanner to ensure real-world applicability. This structured comparison allowed for a direct assessment of image quality improvements and distortion reduction.

Main Results:

Key findings from the literature reveal that the 3D OGprep-GRASE sequence reduces scan time by a factor of 1.38. The method increases signal-to-noise ratio by 1.74-2.27 times compared to 2D-EPI for thick slice coverage. This enhanced signal strength enables more precise diffusion-tensor reconstruction in multishot protocols. Furthermore, the GRASE-based approach significantly lowers image distortion compared to conventional 2D-EPI. Diffusivity measurements consistently demonstrate time-dependency in both white and gray matter tissues. These results hold true across both the novel and traditional imaging sequences tested. The data confirm that the 3D preparation module maintains sensitivity to microstructural changes during brain scans. Overall, the sequence provides a more efficient and accurate alternative for clinical diffusion imaging.

Conclusions:

The proposed three-dimensional sequence offers a viable path for translating time-dependent diffusion measurements into routine clinical practice. Authors report that this approach effectively minimizes image distortion while simultaneously boosting signal strength. Synthesis and implications suggest that the method provides a robust alternative to standard two-dimensional multislice protocols. Researchers observed that the technique maintains consistent sensitivity to tissue microstructure across different brain regions. Findings indicate that the improved signal-to-noise ratio facilitates more accurate reconstruction of diffusion tensors. The study demonstrates that clinical scanners can achieve high-quality results without excessive scan times. These improvements may enhance the utility of diffusion imaging for characterizing complex brain architecture. Future applications might leverage this increased efficiency to expand the scope of non-invasive neurological assessments.

The researchers propose a 3D OGprep-GRASE sequence that integrates global saturation, diffusion encoding, and GRASE readout modules. This configuration achieves a 1.38-fold reduction in scan time and increases signal-to-noise ratio by 1.74 to 2.27 times compared to traditional 2D-EPI methods.

The authors utilize multiplexed sensitivity-encoding reconstruction to address phase errors arising from the multishot acquisition process. This computational tool is necessary to ensure image integrity when combining data from multiple segments, unlike standard single-shot approaches that do not require such complex phase correction.

A 3T clinical system is necessary to implement this sequence, as the authors designed the protocol specifically for standard hospital hardware. This field strength provides the baseline signal required to support the 3D GRASE readout while maintaining the sensitivity needed for oscillating gradient diffusion measurements.

The researchers employ 3D GRASE readout modules to replace conventional 2D-EPI. This component role is to reduce geometric distortion and increase efficiency, whereas 2D-EPI is prone to significant susceptibility artifacts that degrade image quality in thick slice coverage scenarios.

The team measured diffusivity at 50 Hz and 25 Hz for oscillating gradients, alongside 30 ms and 60 ms for pulsed-gradient diffusion. These measurements reveal clear time-dependent changes in both white and gray matter, confirming the technique captures microstructural variations accurately.

The authors propose that their method facilitates the clinical translation of time-dependent diffusion MRI. By overcoming existing speed and quality barriers, they suggest this approach makes advanced microstructural mapping more feasible for routine patient examinations on standard clinical platforms.