Radiation Eye Dose for Physicians in CT Fluoroscopy-Guided Biopsy.
Yohei Inaba1,2, Shin Hitachi3, Munenori Watanuki4
1Course of Radiological Technology, Health Sciences, Tohoku University Graduate School of Medicine, 2-1 Seiryo, Aoba, Sendai 980-8575, Miyagi, Japan.
Summary
Physicians performing computed tomography fluoroscopy (CTF)-guided biopsies receive significant radiation eye doses. Direct eye dosimeters like DOSIRIS are crucial for accurate measurement, as neck dosimeters underestimate eye dose (Hp (3)).
Area of Science:
- Medical Physics
- Radiology
- Occupational Health
Background:
- Physicians performing computed tomography fluoroscopy (CTF)-guided biopsies are at risk of elevated radiation eye doses due to proximity to scattered radiation.
- Accurate dosimetry is essential for monitoring and mitigating occupational radiation exposure to the lens of the eye.
Purpose of the Study:
- To measure and evaluate the radiation eye dose (Hp (3)) received by a physician during CTF-guided biopsies over 18 months.
- To assess the protective efficacy of lead (Pb) eyeglasses and compare direct eye dosimetry with neck dosimetry and CT dose parameters.
Main Methods:
- A single physician wore a direct eye dosimeter (DOSIRIS™) above and below lead eyeglasses during 95 CTF-guided biopsy procedures.
- Occupational radiation dose was also monitored using a neck dosimeter, and CT dose-related parameters were recorded.
- Eye doses were estimated using neck dosimeters and CT parameters for comparison.
Main Results:
- The physician's 18-month eye doses (Hp (3)) without lead eyeglasses were 2.25 mSv (right) and 2.06 mSv (left).
- Lead eyeglasses provided 27.8% (right) and 37.5% (left) protection.
- Significant correlations were found between eye and neck doses (R²=0.82/0.55) and between CT parameters (mAs) and eye doses (R²=0.50/0.52).
- Neck dosimeters underestimated the actual eye dose (Hp (3)).
Conclusions:
- Direct eye dosimetry, such as with DOSIRIS™, is necessary for accurate assessment of physician eye lens doses during CTF procedures.
- Lead eyeglasses offer significant, though variable, protection against radiation exposure to the eyes.
- Standard occupational monitoring using neck dosimeters is insufficient for evaluating radiation eye dose in CTF settings.