Bioelectric Analyses of an Osseointegrated Intelligent Implant Design System for Amputees
Published on: July 15, 2009
Toward Self-Powered Load Imbalance Detection for Instrumented Knee Implants Using Quadrant Triboelectric Energy
Mahmood Chahari1, Hosein Haghshenas2, Emre Salman2
1State University of New York at Binghamton, Binghamton, NY, USA.
This study introduces a self-powered triboelectric nanogenerator (TENG) pressure sensor for detecting tibial load imbalance in knee implants. Early detection of load shifts can prevent implant loosening, a common cause of failure.
Area of Science:
- Biomedical Engineering
- Materials Science
- Mechanical Engineering
Background:
- Knee implant loosening is a major cause of implant failure.
- Accurate monitoring of load distribution on the tibial tray is crucial for implant longevity.
- Existing monitoring methods may lack self-powering capabilities or real-time feedback.
Purpose of the Study:
- To develop and evaluate a triboelectric nanogenerator (TENG) as a self-powered pressure sensor for measuring load imbalance on the tibial tray.
- To investigate the potential of TENGs for early detection of knee implant loosening.
- To assess the power harvesting capabilities of the TENG system.
Main Methods:
- A segmented quadrant design TENG pressure sensor with various micro-patterns (pyramid, cylindrical, bar) was proposed.
- Experimental testing was conducted to evaluate sensing capabilities and power harvesting under axial forces.
- The relationship between quadrant output and axial force was determined to characterize sensor performance and find the center of pressure.
Main Results:
- The TENG pressure sensors demonstrated varying sensitivity across different quadrants, enabling axial force measurement.
- A clear relationship between each quadrant's output and the applied axial force was established.
- The system successfully characterized self-powered sensor performance and enabled the determination of the center of pressure.
- Under a 2200 N walking load, each quadrant produced measurable apparent power at 1 Hz and 2 Hz.
Conclusions:
- The developed TENG-based pressure sensor is a viable self-powered solution for monitoring load imbalance on tibial trays.
- Detecting shifts in the center of pressure can serve as an early indicator of knee implant loosening.
- The TENG system exhibits potential for both sensing and energy harvesting in instrumented implants.
More Related Videos
08:08Oscillation and Reaction Board Techniques for Estimating Inertial Properties of a Below-knee Prosthesis
Published on: May 8, 2014
11:16Engineering Platform and Experimental Protocol for Design and Evaluation of a Neurally-controlled Powered Transfemoral Prosthesis
Published on: July 22, 2014
Related Concept Videos
Ammeter
When an ammeter is used to measure the...
Voltmeter
An ideal voltmeter would have infinite resistance, so connecting it between two points in a circuit would not alter any of the currents. Real voltmeters always have finite...
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
Three-Phase Short Circuit—Unloaded Synchronous Machine
This behavior occurs due to the magnetic flux produced by the short-circuit armature currents. Initially, these currents follow high-reluctance paths but eventually shift to...
Electronic Distance Measuring Instruments